Specifications – Atec Mitutoyo-SJ-410 User Manual

Page 11

background image

11

Specifications

Model No.

SJ-411

SJ-412

inch/mm

178-581-01A

178-581-02A

178-583-01A

178-583-02A

Measuring

range

X axis

25mm (1inch)

50mm (2inch)

Z1 axis (detector unit)

800µm, 80µm, 8µm

*Up to 2,400µm with an optional stylus

Detector

Measuring principle

Differential inductance

Resolution

0.01µm (800µm range) / 0.001µm (80µm range) / 0.0001µm (8µm range)

0.4µinch (32000µinch) / 0.04µinch (3200µinch) / 0.004µinch (320µinch)

Stylus tip

60°/2µm (80µinch)

90°/5µm (200µinch)

60°/2µm (80µinch)

90°/5µm (200µinch)

Measuring force

0.75mN

4mN

0.75mN

4mN

Radius of skid curvature

R40 mm (R1.57")

Measuring method

Skidded measurement / skidless measurement

Drive unit: X-axis

Measuring speed

0.05, 0.1, 0.2, 0.5, 1.0mm/s (0.002, 0.004, 0.02, 0.04 inch/s)

Drive speed

0.5, 1, 2, 5mm/s (0.02, 0.04, 0.08, 0.2 inch/s)

Straightness

0.3µm / 25mm (12µinch/ 1inch)

0.5µm / 50mm (20µinch/ 2inch)

Height-tilt

adjustment unit

Height adjustment

10mm (0.39inch)

Tilt adjustment

±1.5°

Standards

JIS1982 / JIS1994 / JIS2001 / ISO1997 / ANSI / VDA

Parameters

Ra, Rq, Rz, Ry, Rp, Rv, Rt, R3z, Rsk, Rku, Rc, RPc, RSm, Rmax*

1

, Rz1max*

2

, S, HSC, RzJIS*

3

, Rppi, RΔa, RΔq, Rlr, Rmr, Rmr(c),

Rσc, Rk, Rpk, Rvk, Mr1, Mr2, A1, A2, Vo, λa, λq, Lo, Rpm, tp*

4

, Htp*

4

, R, Rx, AR, W, AW, Wx, Wte, Possible Customize

Measured profiles

Primary, Roughness, DF, Filtered waviness curve, R-Motif, W-Motif

Graph analysis

BAC and ADC curves

Data compensation

Parabola/ Hyperbola/ Ellipse/ Circle/ Conic/ Tilting, Compensation off

Filter

2CR, PC75, Gaussian filter

Cut-off length

λc

0.08, 0.25, 0.8, 2.5, 8.0mm

λs *

5

2.5, 8.0, 25mm (100, 320, 1000µinch)

Sample length

0.08, 0.25, 0.8, 2.5, 8.0, 25.0mm

Number of sampling lengths

Ч1, Ч2, Ч3, Ч4, Ч5, Ч6, Ч7, Ч8, Ч9, Ч10, Ч11, Ч12, Ч13, Ч14, Ч15, Ч16, Ч17, Ч18, Ч19, Ч20

Arbitrary length

0.1~25mm

0.1~50mm

Functions

Customization

Desired parameters can be selected for calculation and display

Simple contour analysis function

Step, Step volume, Dimensions, Coordinate difference

DAT function

Helps to adjust leveling during skidless measurement

Real sampling function

Samples stylus displacement for a specified time without engaging detector traverse.

Statistical processing

Static measurement (max. 3 parameters) is possible. Static processing for MAX, MIN, AVERAGE,

standard deviation, histogram and pass rate is possible

GO/ NG judgement*

6

Max rule / 16% rule / Average rule / Standard deviation (1σ, 2σ, 3σ)

Storage functions

10 measuring conditions can be stored in internal memory

Printing function

Measurement conditions / Calculation results / GO / NG judgement result / Calculation results for each sampling length /

Measurement curve / BAC / ADC / Environmental setting information

Display languages

Japanese, English, German, French, Italian, Spanish, Portuguese, Korean,

Traditional Chinese, Simplified Chinese, Czech, Polish, Hungarian Turkish, Swedish, Dutch

Storage

Internal memory: Measurement condition (10 sets)

Memory card (option): 500 measurement condition, 10000 measuring data, 10000 text data, 500 statistic data,

1 backup of machine setting, the last ten traces (Trace 10)

External I/O

USB I/F, Digimatic output, RS-232C I/F, External SW I/F

Power supply

Battery

Two-way power supply: battery (rechargeable Ni-MH battery) and AC adapter

*Charging time: about 4 hours (may vary due to ambient temperature)

*Endurance: about 1500 measurements (differs slightly due to use conditions / environment)

Power consumption

50W

Size

(W×D×H)

Display unit

275Ч198Ч109mm (10.83Ч4.29Ч7.80inch)

Height adjustment unit

130.9Ч63Ч99mm (5.16Ч2.48Ч3.90 inch)

Drive unit

128Ч35.8Ч46.6mm (5.04Ч1.41Ч1.83 inch)

154.5Ч35.8Ч46.6mm (6.08Ч1.41Ч1.83inch)

Mass

Display unit

1.7kg

Height adjustment unit

0.4kg

Drive unit

0.6kg

0.64kg

Standard accessories

*1: Only for VDA/ANSI/JIS'82 standards.

*2: Only for JIS'97 standard.

*3: Only for JIS'01 standard.

*4: Only for ANSI standard.

*5: λs may not be switchable depending on standard selected.

*6: Standard deviation only can be selected in ANSI.16% rule cannot be selected in VDA.

*7: Either No.178-396-2 or No.178-397-2 is supplied as a standard accessory depending on the Order No. of the main unit for SJ-410 Series.

*8: The standard stylus (No.12AAC731 or No.12AAB403), which is compatible with the detector supplied, is a standard accessory.

Detector*

7

, Stylus*

8

, Roughness specimen

270732

Printing paper

12BAL402 Touch-screen protection sheet

12BAG834 Touch pen

12AAN041 Carrying case

AC adapter, Philips screwdriver,

Strap for stylus pen, Operation manual,

Quick reference manual, Warranty