Specifications – Atec Mitutoyo-SJ-410 User Manual
Page 11
11
Specifications
Model No.
SJ-411
SJ-412
inch/mm
178-581-01A
178-581-02A
178-583-01A
178-583-02A
Measuring
range
X axis
25mm (1inch)
50mm (2inch)
Z1 axis (detector unit)
800µm, 80µm, 8µm
*Up to 2,400µm with an optional stylus
Detector
Measuring principle
Differential inductance
Resolution
0.01µm (800µm range) / 0.001µm (80µm range) / 0.0001µm (8µm range)
0.4µinch (32000µinch) / 0.04µinch (3200µinch) / 0.004µinch (320µinch)
Stylus tip
60°/2µm (80µinch)
90°/5µm (200µinch)
60°/2µm (80µinch)
90°/5µm (200µinch)
Measuring force
0.75mN
4mN
0.75mN
4mN
Radius of skid curvature
R40 mm (R1.57")
Measuring method
Skidded measurement / skidless measurement
Drive unit: X-axis
Measuring speed
0.05, 0.1, 0.2, 0.5, 1.0mm/s (0.002, 0.004, 0.02, 0.04 inch/s)
Drive speed
0.5, 1, 2, 5mm/s (0.02, 0.04, 0.08, 0.2 inch/s)
Straightness
0.3µm / 25mm (12µinch/ 1inch)
0.5µm / 50mm (20µinch/ 2inch)
Height-tilt
adjustment unit
Height adjustment
10mm (0.39inch)
Tilt adjustment
±1.5°
Standards
JIS1982 / JIS1994 / JIS2001 / ISO1997 / ANSI / VDA
Parameters
Ra, Rq, Rz, Ry, Rp, Rv, Rt, R3z, Rsk, Rku, Rc, RPc, RSm, Rmax*
1
, Rz1max*
2
, S, HSC, RzJIS*
3
, Rppi, RΔa, RΔq, Rlr, Rmr, Rmr(c),
Rσc, Rk, Rpk, Rvk, Mr1, Mr2, A1, A2, Vo, λa, λq, Lo, Rpm, tp*
4
, Htp*
4
, R, Rx, AR, W, AW, Wx, Wte, Possible Customize
Measured profiles
Primary, Roughness, DF, Filtered waviness curve, R-Motif, W-Motif
Graph analysis
BAC and ADC curves
Data compensation
Parabola/ Hyperbola/ Ellipse/ Circle/ Conic/ Tilting, Compensation off
Filter
2CR, PC75, Gaussian filter
Cut-off length
λc
0.08, 0.25, 0.8, 2.5, 8.0mm
λs *
5
2.5, 8.0, 25mm (100, 320, 1000µinch)
Sample length
0.08, 0.25, 0.8, 2.5, 8.0, 25.0mm
Number of sampling lengths
Ч1, Ч2, Ч3, Ч4, Ч5, Ч6, Ч7, Ч8, Ч9, Ч10, Ч11, Ч12, Ч13, Ч14, Ч15, Ч16, Ч17, Ч18, Ч19, Ч20
Arbitrary length
0.1~25mm
0.1~50mm
Functions
Customization
Desired parameters can be selected for calculation and display
Simple contour analysis function
Step, Step volume, Dimensions, Coordinate difference
DAT function
Helps to adjust leveling during skidless measurement
Real sampling function
Samples stylus displacement for a specified time without engaging detector traverse.
Statistical processing
Static measurement (max. 3 parameters) is possible. Static processing for MAX, MIN, AVERAGE,
standard deviation, histogram and pass rate is possible
GO/ NG judgement*
6
Max rule / 16% rule / Average rule / Standard deviation (1σ, 2σ, 3σ)
Storage functions
10 measuring conditions can be stored in internal memory
Printing function
Measurement conditions / Calculation results / GO / NG judgement result / Calculation results for each sampling length /
Measurement curve / BAC / ADC / Environmental setting information
Display languages
Japanese, English, German, French, Italian, Spanish, Portuguese, Korean,
Traditional Chinese, Simplified Chinese, Czech, Polish, Hungarian Turkish, Swedish, Dutch
Storage
Internal memory: Measurement condition (10 sets)
Memory card (option): 500 measurement condition, 10000 measuring data, 10000 text data, 500 statistic data,
1 backup of machine setting, the last ten traces (Trace 10)
External I/O
USB I/F, Digimatic output, RS-232C I/F, External SW I/F
Power supply
Battery
Two-way power supply: battery (rechargeable Ni-MH battery) and AC adapter
*Charging time: about 4 hours (may vary due to ambient temperature)
*Endurance: about 1500 measurements (differs slightly due to use conditions / environment)
Power consumption
50W
Size
(W×D×H)
Display unit
275Ч198Ч109mm (10.83Ч4.29Ч7.80inch)
Height adjustment unit
130.9Ч63Ч99mm (5.16Ч2.48Ч3.90 inch)
Drive unit
128Ч35.8Ч46.6mm (5.04Ч1.41Ч1.83 inch)
154.5Ч35.8Ч46.6mm (6.08Ч1.41Ч1.83inch)
Mass
Display unit
1.7kg
Height adjustment unit
0.4kg
Drive unit
0.6kg
0.64kg
Standard accessories
*1: Only for VDA/ANSI/JIS'82 standards.
*2: Only for JIS'97 standard.
*3: Only for JIS'01 standard.
*4: Only for ANSI standard.
*5: λs may not be switchable depending on standard selected.
*6: Standard deviation only can be selected in ANSI.16% rule cannot be selected in VDA.
*7: Either No.178-396-2 or No.178-397-2 is supplied as a standard accessory depending on the Order No. of the main unit for SJ-410 Series.
*8: The standard stylus (No.12AAC731 or No.12AAB403), which is compatible with the detector supplied, is a standard accessory.
Detector*
7
, Stylus*
8
, Roughness specimen
270732
Printing paper
12BAL402 Touch-screen protection sheet
12BAG834 Touch pen
12AAN041 Carrying case
AC adapter, Philips screwdriver,
Strap for stylus pen, Operation manual,
Quick reference manual, Warranty