Electrostatic Discharge Simulator
5
ESD Simulator ESS-2002EX
&
&
TC-815R
ESD Simulator
Conforming to IEC61000-4-2
A manual ESD simulator that is both lightweight and
compact. A completely new design has made the product
easier to use, safer and more reliable.
FEATURES
Easy to use self-explanatory control panel
Preset discharge interval and count
Preset IEC standard test levels
“V STEP” function enables the voltage setting can be altered
in steps of the desired preset increment (or decrement)
■
•
•
•
•
CONTROLS INDICATORS AND DISPLAYS
■
Specifi cations
■
Item
Specifi cations
Output polarity
Positive or negative
Output voltage
0.20kV ∼ 30.0kV± 5% (30.5kV max)
0.20kV ∼ 1.0kV: 0.05kV step, 1.0kV ∼ 30.0kV:0.1kV step
Discharge interval
(contact discharge)
Normal mode: 0.05s ∼ 9.99s± 10%
Extra mode : 1.00s ∼ 9.99s± 10%
0.01s step
No. of times of discharge
1 ∼ 999 times preset on the coutner or continuous
Discharge mode
Air/contact discharge
Radiation level
Normal/Extra mode
Trigger mode
Gun trigger or main unit trigger
Preset IEC levels
Contact discharge: 2.0kV, 4.0kV, 6.0kV, 8.0kV (4 test voltages preset)
Air discharge: 2.0kV, 4.0kV, 8.0kV, 15.0kV (4 test votlages preset)
Preset step voltage mode
(“V STEP” function)
Step setting: 0.00kV ∼ 10kV
0.2kV ∼ 1.0kV: 0.05kV step, 1.0kV ∼ 30.0kV: 0.1kV step
Each time Increment or Decrement key is pressed, the test voltage changes by the preset step
Warning lamp
Lights up while the HV circuitry is on.
Blinks while discharges are in progress
Discharge gun
TC-815R
Charging resistance
10MΩ(53 MΩ for combination with TC-815R discharge gun)
Power supply
100 ∼ 240V AC 50/60 Hz ± 10%
Power consumption
62VA
Operating temperature and humidity
+ 15 ∼ +35℃ 25 ∼ 75% (no condensatoin)
Dimensions and Weight
340 (W) X 200 (H) X 170 (D) mm (projection excluded) Approx 4.6 kg
+/- Selects the polarity
Displays output voltage, discharge interval and number of discharges settings.
Increment and decrement keys: change the value for the selected parameter
CONTACT/AIR selects either air or contact discharge
V STEP: when this enabled, each time Increment or Decrement key is pressed, the test voltage changes by
the preset step.
TRIGGER: selects either CONTROLER (TRIGGER key on ESS main unit) or TC-815R gun as the operative trigger
Warning lamp: alight while the HV circuitry is on and blinks while discharges are in progress
STOP key: turns off the HV circuitry
START key: turns on the HV circuitry
SELECT key: selects the parameter to be changed. Each time the key is pressed, VOLTAGE, INTERVAL and COUNT
indicator starts blinking in turn. While blinking, the value can be changed by using the Increment or Decrement key
IEC LEVEL: selects either of the preset IEC specified test voltages. The preset voltages are 2.0kV, 4kV, 6kV
and 8kV for contact discharge and 2.0kV, 4.0kV, 8.0kV and 15.0kV for air discharge as per IEC61000-4-2
EXTRA/NORMAL: provides the selection of NORMAL or EXTRA in terms of the strength of radiation from the
discharge gun. The EXTRA mode reduces the radiation. In the EXTRA mode, faster than 1.0s setting for
discharge interval is not accepted.
ESS-2002EX & TC-815R (with PS-806 gun stand)