Table 60503-2. Programming Ranges (continued)

Function

Front Panel

Front Panel

HPSL Command

Range of Values

Key

Display

(Short Form)

Transient Operation

Set Frequency

FREQ value

"TRAN:FREQ value"

0.25 Hz to 10 kHz

Set Duty Cycle

(shift)

DCYCLE value

"TRAN:DCYC value"

3-97% (0.25 Hz-1 kHz)

6-94% (1 kHz-10 kHz)

*Set Pulse Width

"TRAN:TWID value"

0.00005 to 4 s

Trigger Operation

*Set Trigger Period

"TRIG:TIM value"

0.000008 to 4 s

Current Protection

*Set Current Level

"CURR:PROT value"

0 to 10.2 A

*Set Delay Time

"CURR:PROT:DEL value"

0 to 60 s

*Can only be programmed remotely via the GPIB.

Table 60503-3. Factory Default Settings

Function

Settings

Function

Setting

CURR level

0 A

Mode (CC, CR, CV)

CC

CURR transient level

0 A

Input (on/off)

on

*CURR slew rate

0.17 A/

µ

s

Short (on/off)

off

CURR range

10 A

Transient operation (on/off)

off

*CURR protection (on/off)

off

***TRAN mode

continuous

**CURR protection level

10.2 A

(continuous, pulse, toggle)

**CURR protection delay

15 s

TRAN frequency

1 kHz

TRAN duty cycle

50%

RES level

50 k

**TRAN pulse width

0.5 ms

RES transient level

50 k

RES range

50 k

**TRIG source

hold

(bus, external, hold, timer, line)

VOLT level

240 V

**TRIG period

0.001 s

VOLT transient level

240 V

**PORT0 output (on/off)

off (logic 0)

VOLT slew rate

2 V/

µ

s

**CAL mode (on/off)

off

The *RST command resets the CURR slew rate to 0.83 A/

µ

, not to the factory default.

**Can only be programmed remotely via the GPIB.
***Continuous transient mode is the only mode available at the front panel. Pulsed, toggled, and continuous modes can
all be programmed remotely via the GPIB.