Atec Agilent-60503B User Manual
Page 2
![background image](/manuals/318501/2/background.png)
3
Table 60503-1 Specifications and Supplemental Characteristics (continued)
Transient Operation:
Continuous Mode
Frequency Range:
0.25 Hz to 10 kHz
Frequency Resolution:
4%
Frequency Accuracy:
3%
Duty Cycle Range:
3% to 97% (0.25 Hz to 1 kHz); 6% to 94% (1 kHz to 10 kHz)
Duty Cycle Resolution:
4%
Duty Cycle Accuracy:
6% of setting ± 2%
Pulsed Mode
Pulse Width:
50
µ
s ± 3% minimum; 4 s ± 3% maximum
Transient Current Level (0 to 1 A and 0 to 10 A ranges):
Resolution:
4 mA (1 A range); 43 mA (10 A range)
Accuracy:
± 0.18% ± 13 mA (1 A range); ± 0.18% ± 50 mA (10 A range)
Temperature Coefficient:
180 ppm/
°
C ± 1.2 mA/
°
C
Transient Resistance Level (0.20 to 24
Ω
, 24
Ω
to 10 k
Ω
, and 240
Ω
to 50 k
Ω
ranges):
Resolution:
100 m
Ω
(24
Ω
range); 0.18 mS (10 k
Ω
range); 0.018 mS (50 k
Ω
range)
Accuracy:
± 0.8% + 200 m
Ω
with > 1 A at input (24
Ω
range)
± 0. 3% + 0.5 mS with
≥
24 V at input (10 k
Ω
range)
± 0.3% + 0.4 mS with
≥
24 V at input (50 k
Ω
range)
Transient Voltage Level (0 to 240 V):
Resolution:
1.0
V
Accuracy:
± 0.15% ± 1.1 V
Temperature Coefficient:
120
ppm/
°
C ± 10 mV/
°
C
Current Readback:
Resolution:
2.7 mA (via GPIB); 10 mA (front panel)
Accuracy:
(after 30 minute wait): ± 0.12% ± 10 mA
Temperature Coefficient:
100
ppm/
°
C ± l mA/
°
C
Voltage Readback:
Resolution:
67 mV (via GPIB); 100 mV (front panel)
Accuracy:
± 0.1% ± 150 mV
Temperature Coefficient:
100 ppm/
°
C ± 8 mV/
°
C
Maximum Readback Capability:
260 V (typical)
Power Readback:
Accuracy:
± 0.2% ± 3 W