Table 60503-1 Specifications and Supplemental Characteristics (continued)

Transient Operation:

Continuous Mode

Frequency Range:

0.25 Hz to 10 kHz

Frequency Resolution:

4%

Frequency Accuracy:

3%

Duty Cycle Range:

3% to 97% (0.25 Hz to 1 kHz); 6% to 94% (1 kHz to 10 kHz)

Duty Cycle Resolution:

4%

Duty Cycle Accuracy:

6% of setting ± 2%

Pulsed Mode
Pulse Width:

50

µ

s ± 3% minimum; 4 s ± 3% maximum

Transient Current Level (0 to 1 A and 0 to 10 A ranges):

Resolution:

4 mA (1 A range); 43 mA (10 A range)

Accuracy:

± 0.18% ± 13 mA (1 A range); ± 0.18% ± 50 mA (10 A range)

Temperature Coefficient:

180 ppm/

°

C ± 1.2 mA/

°

C

Transient Resistance Level (0.20 to 24

, 24

to 10 k

, and 240

to 50 k

ranges):

Resolution:

100 m

(24

range); 0.18 mS (10 k

range); 0.018 mS (50 k

range)

Accuracy:

± 0.8% + 200 m

with > 1 A at input (24

range)

± 0. 3% + 0.5 mS with

24 V at input (10 k

range)

± 0.3% + 0.4 mS with

24 V at input (50 k

range)

Transient Voltage Level (0 to 240 V):

Resolution:

1.0

V

Accuracy:

± 0.15% ± 1.1 V

Temperature Coefficient:

120

ppm/

°

C ± 10 mV/

°

C

Current Readback:

Resolution:

2.7 mA (via GPIB); 10 mA (front panel)

Accuracy:

(after 30 minute wait): ± 0.12% ± 10 mA

Temperature Coefficient:

100

ppm/

°

C ± l mA/

°

C

Voltage Readback:

Resolution:

67 mV (via GPIB); 100 mV (front panel)

Accuracy:

± 0.1% ± 150 mV

Temperature Coefficient:

100 ppm/

°

C ± 8 mV/

°

C

Maximum Readback Capability:

260 V (typical)

Power Readback:

Accuracy:

± 0.2% ± 3 W