Atec Agilent-60503B User Manual
Page 4
5
Table 60503-1 Specifications and Supplemental Characteristics (continued)
Current Slew Rate:*
Rate #
10 A Range Step
1 A Range Step
Transition Time
1
0.17 A/ms
17 A/s
8.0 ms
2
0.42 A/ms
42 A/s
3.2 ms
3
0.83 A/ms
83 A/s
1.6 ms
4
1.7 A/ms
0.17 A/ms
800
µ
s
5
4.2 A/ms
0.42 A/ms
320
µ
s
6
8.3 A/ms
0.83 A/ms
160
µ
s
7
17 A/ms
1.7 A/ms
80
µ
s
8
42 A/ms
4.2 A/ms
32
µ
s
9
83 A/ms
8.3 A/ms
20
µ
s
10
0.17 A/
µ
s
17 A/ms
20
µ
s
11
0.42 A/
µ
s
42 A/ms
16
µ
s
12
0.83 A/
µ
s
83 A/ms
16
µ
s
*AC performance specified from 3 to 240 V.
Voltage Slew Rate:
Rate #
Voltage Range Step
Transition Time*
1
4 V/ms
8.0 ms
2
10 V/ms
3.2 ms
3
20 V/ms
1.6 ms
4
40 V/ms
800
µ
s
5
100 V/ms
320
µ
s
6
200 V/ms
160
µ
s
7
0.4 V/
µ
s
100
µ
s
8
1 V/
µ
s
100
µ
S
9
2 V/
µ
s
100
µ
S
*Transition time based on low capacitance current source.
Resistance Slew Rate
(24
Ω
range): Uses the value programmed for voltage slew rate.
Resistance Slew Rate
(10 k and 50 k
Ω
ranges): Uses the value programmed for current slew rate.
Transient Current Overshoot (When programmed from 0A):
Range
Transient Current Level
Current Slew Rate
Overshoot*
10 A
2-10 A
All slew rates
0
0.5 A
0.17 A/
µ
s to 0.83 A/
µ
s
5%
0.5 A
0.17 A/ms to 42 A/ms
0
1 A
0.83 A/
µ
s
1%
1 A
0.17 A/ms to 0.17 A/
µ
s
0
1 A
0.5 A
8.3 A/ms
4%
0.5 A
0.17 A/s and 0.17 A/ms
0
1 A
All slew rates
0
*All overshoot values assume a total inductance of l
µ
H, or less, in the load leads connected to the D.U.T.