Table 60503-1 Specifications and Supplemental Characteristics (continued)

Current Slew Rate:*

Rate #

10 A Range Step

1 A Range Step

Transition Time

1

0.17 A/ms

17 A/s

8.0 ms

2

0.42 A/ms

42 A/s

3.2 ms

3

0.83 A/ms

83 A/s

1.6 ms

4

1.7 A/ms

0.17 A/ms

800

µ

s

5

4.2 A/ms

0.42 A/ms

320

µ

s

6

8.3 A/ms

0.83 A/ms

160

µ

s

7

17 A/ms

1.7 A/ms

80

µ

s

8

42 A/ms

4.2 A/ms

32

µ

s

9

83 A/ms

8.3 A/ms

20

µ

s

10

0.17 A/

µ

s

17 A/ms

20

µ

s

11

0.42 A/

µ

s

42 A/ms

16

µ

s

12

0.83 A/

µ

s

83 A/ms

16

µ

s

*AC performance specified from 3 to 240 V.

Voltage Slew Rate:

Rate #

Voltage Range Step

Transition Time*

1

4 V/ms

8.0 ms

2

10 V/ms

3.2 ms

3

20 V/ms

1.6 ms

4

40 V/ms

800

µ

s

5

100 V/ms

320

µ

s

6

200 V/ms

160

µ

s

7

0.4 V/

µ

s

100

µ

s

8

1 V/

µ

s

100

µ

S

9

2 V/

µ

s

100

µ

S

*Transition time based on low capacitance current source.

Resistance Slew Rate

(24

range): Uses the value programmed for voltage slew rate.

Resistance Slew Rate

(10 k and 50 k

ranges): Uses the value programmed for current slew rate.

Transient Current Overshoot (When programmed from 0A):

Range

Transient Current Level

Current Slew Rate

Overshoot*

10 A

2-10 A

All slew rates

0

0.5 A

0.17 A/

µ

s to 0.83 A/

µ

s

5%

0.5 A

0.17 A/ms to 42 A/ms

0

1 A

0.83 A/

µ

s

1%

1 A

0.17 A/ms to 0.17 A/

µ

s

0

1 A

0.5 A

8.3 A/ms

4%

0.5 A

0.17 A/s and 0.17 A/ms

0

1 A

All slew rates

0

*All overshoot values assume a total inductance of l