beautypg.com

Atec Agilent-60503B User Manual

Page 3

background image

4

Table 60503-1 Specifications and Supplemental Characteristics (continued)

External Analog Programming 0 to 10 V (dc or ac):

Bandwidth:

10 kHz (3 db frequency)

Accuracy:

± 3% ± 10 mA (0 to 1 A range)
± 3% ± 20 mA (0 to 10 A range)
± 0.5% ± 150 mV (0 to 240 V range)

Temperature Coefficient:

150 ppm/

°

C ± 1 mA/

°

C (current ranges)

120 ppm/

°

C ± 10 mV/

°

C (voltage range)

External Current Monitor (0 to 10 V):

Accuracy:

± 3% ± 10 mA (referenced to analog common)

Temperature Coefficient:

100 ppm/

°

C ± l mA/

°

C

External Voltage Monitor

(0 to 10 V):

Accuracy:

± 0.4% ± 240 mV (referenced to analog common)

Temperature Coefficient:

70

ppm/

°

C ± 1.2 mV/

°

C

Remote Sensing:

5 Vdc maximum between sense and input binding posts

Maximum Input Levels:

Current:

10.2 A (programmable to lower limits)

Voltage:

250

V

Minimum Operating Voltage:

2 V (derated to 0 V at 0 A)

PARD (20 Hz to 10 MHz noise):

Current:

1 mA rms/10 mA p-p

Voltage:

6 mV rms

DC Isolation Voltage:

± 240 Vdc between + or - input binding post and chassis ground

Digital Inputs:

Vlo:

0.9 V maximum at Ilo = -1 mA

Vhi

3.15 V minimum (pull-up resistor on input)

Digital Outputs:

V

lo:

0.72 V maximum at Ilo = 1 mA

Vhi:

4.4 V minimum at Ilo - 20

µ

A

SUPPLEMENTAL CHARACTERISTICS

Programmable Slew Rate (For any given input transition, the time required will be either the total slew time or a minimum
transition time, whichever is longer. The minimum transition time increases when operating with input currents under 0.2 A
and decreases with input currents over 2 A. The following are typical values; ± 25% tolerance):