Atec Agilent-60503B User Manual
Page 3
4
Table 60503-1 Specifications and Supplemental Characteristics (continued)
External Analog Programming 0 to 10 V (dc or ac):
Bandwidth:
10 kHz (3 db frequency)
Accuracy:
± 3% ± 10 mA (0 to 1 A range)
± 3% ± 20 mA (0 to 10 A range)
± 0.5% ± 150 mV (0 to 240 V range)
Temperature Coefficient:
150 ppm/
°
C ± 1 mA/
°
C (current ranges)
120 ppm/
°
C ± 10 mV/
°
C (voltage range)
External Current Monitor (0 to 10 V):
Accuracy:
± 3% ± 10 mA (referenced to analog common)
Temperature Coefficient:
100 ppm/
°
C ± l mA/
°
C
External Voltage Monitor
(0 to 10 V):
Accuracy:
± 0.4% ± 240 mV (referenced to analog common)
Temperature Coefficient:
70
ppm/
°
C ± 1.2 mV/
°
C
Remote Sensing:
5 Vdc maximum between sense and input binding posts
Maximum Input Levels:
Current:
10.2 A (programmable to lower limits)
Voltage:
250
V
Minimum Operating Voltage:
2 V (derated to 0 V at 0 A)
PARD (20 Hz to 10 MHz noise):
Current:
1 mA rms/10 mA p-p
Voltage:
6 mV rms
DC Isolation Voltage:
± 240 Vdc between + or - input binding post and chassis ground
Digital Inputs:
Vlo:
0.9 V maximum at Ilo = -1 mA
Vhi
3.15 V minimum (pull-up resistor on input)
Digital Outputs:
V
lo:
0.72 V maximum at Ilo = 1 mA
Vhi:
4.4 V minimum at Ilo - 20
µ
A
SUPPLEMENTAL CHARACTERISTICS
Programmable Slew Rate (For any given input transition, the time required will be either the total slew time or a minimum
transition time, whichever is longer. The minimum transition time increases when operating with input currents under 0.2 A
and decreases with input currents over 2 A. The following are typical values; ± 25% tolerance):