Specifi cations, continued – Atec Agilent-6652A User Manual
Page 3
3
Specifi cations, continued
Specifications
(at 0 ° to 55 °C unless otherwise specified)
6651A-J03
Special order
option
6651A-J09
Special order
option
6652A-J03
Special order
option
6653A-J04
Special order
option
6653A-J17
Special order
option
Number of outputs
1 1 1 1 1
GPIB
Yes Yes Yes Yes Yes
Output ratings
Output voltage
6 V
17 V/20 V
27 V
40 V
30 V
Output current (40 °C)
60 A
30 A/15 A
18.5 A
12.5 A
17.5 A
Maximum current (50 °C/55 °C)
54 A/51 A
27 A/25.5 A
13.5 A/12.75 A
16.65 A/15.72 A 11.25 A/10.6 A
15.75 A/14.87 A
Programming accuracy (at 25 °C ± 5 °C)
Voltage
0.06% +
5 mV
10 mV
13.5 mV
17.5 mV
15 mV
Current
0.15% +
75 mA
36 mA
25 mA
13 mA
16 mA
Ripple and noise from 20 Hz to 20 MHz
Voltage
rms
300 μV
300 μV
450 μV
1.6 mV
400 μV
peak-to-peak
3 mV
4 mV
4.5 mV
5 mV
4 mV
Current
rms
30 mA
13 mA
10 mA
5 mA
6 mA
Readback accuracy at 25 °C ± 5 °C
(percent of reading plus fi xed)
Voltage
0.07% +
6 mV
15 mV
20.5 mV
30 mV
25 mV
+Current
0.15% +
80 mA
40 mA
26 mA
15 mA
18 mA
–Current
0.35% +
150 mA
55 mA
44 mA
24 mA
28 mA
Load regulation
Voltage
1 mV
2 mV
2 mV
3.5 mV
3 mV
Current
6.5 mA
2 mA
1 mA
1 mA
0.5 mA
Line regulation
Voltage
0.5 mV
0.5 mV
0.5 mV
1 mV
1 mV
Current
2 mA
2 mA
2 mA
0.75 mA
0.75 mA
Transient response time
Less than 100 μs for the output voltage to recover to its previous level (within 0.1%
of the voltage rating of the supply or 20 mV, whichever is greater) following any step
change in load current of up to 50% of the output current rating of the supply
Supplemental characteristics
(Non-warranted characteristics determined
by design and useful in applying the product)
6651A-J03
Special order
option
6651A-J09
Special order
option
6652A-J03
Special order
option
6653A-J04
Special order
option
6653A-J17
Special order
option
Average programming resolution
Voltage
2 mV
5 mV
6.75 mV
12 mV
10 mV
Current
18 mA
9 mA
7 mA
4 mA
5 mA
OVP
12 mV
30 mV
30 mV
65 mV
54 mV
OVP accuracy
160 mV
500 mV
400 mV
750 mV
700 mV