Uncertainty calculations for a power measurement (settled, average power)

[Specification values from this document are in bold italic, values calculated on this page are underlined.]

Process:

1. Power level: .............................................................................................................................................. W
2. Frequency: ................................................................................................................................................
3. Calculate meter uncertainty:

Calculate noise contribution

Noise = Noise-per-sample x noise per sample multiplier

Convert noise contribution to a relative term

1

= Noise/Power ...................................................... %

Instrumentation linearity

...................................................................................................................... %

Drift

...........................................................................................................................................................

%

RSS of above three terms ≥ Meter uncertainty = ........................................................................ %

4. Zero uncertainty

(Mode and frequency-dependent) = Zero set/Power = ............................................................. %

5. Sensor calibration uncertainty

(Sensor, frequency, power and temperature-dependent) = ........................................................ %

6. System contribution, coverage factor of 2 ≥ sys

rss

= ........................................................................ %

(RSS three terms from steps 3, 4 and 5)

7. Standard uncertainty of mismatch
Max SWR (frequency-dependent) = .............................................................................................. %

convert to reflection coefficient, | ρ

Sensor

| = (SWR–1)/(SWR+1) = ........................................ %

Max DUT SWR (frequency-dependent) = ...................................................................................... %

convert to reflection coefficient, | ρ

DUT

| = (SWR–1)/(SWR+1) = ........................................... %

8. Combined measurement uncertainty @ k=1

U

C

=

(

Max(ρ

DUT

) • Max(ρ

Sensor

)

)

2

+

(

sys

rss

)

2

.................................................................... %

√2 2

Expanded uncertainty, k = 2, = UC • 2 = ............................................................................................. %

1. The noise-to-power ratio is capped for powers > 100 μW, in these cases use: Noise/100 μW.