Atec Agilent-8990B User Manual
Page 13
13
N1923A/N1924A Wideband Power Sensor Specifications
Sensor
model
Frequency
range
Dynamic range
Rise/fall time
Damage level
Connector
type
N1923A 50 MHz to
18 GHz
–35 dBm to +20 dBm
≤ 3 ns (applicable for
frequencies of ≥ 500
MHz)
+23 dBm (average power);
+30 dBm (< 1 μs duration,
peak power)
Type N (m)
N1924A 50 MHz to
40 GHz
–35 dBm to +20 dBm
≤ 3 ns (applicable for
frequencies of ≥ 500
MHz)
+23 dBm (average power);
+30 dBm (< 1 μs duration,
peak power)
2.4 mm (m)
The N1921A/N1922A P-series wideband power sensors are compatible for use with the 8990B peak power analyzer.
Maximum SWR
Frequency band
N1923A
N1924A
50 MHz to 10 GHz
1.2
1.2
10 GHz to 18 GHz
1.26
1.26
18 GHz to 26.5 GHz
1.3
26.5 GHz to 40 GHz
1.5
Sensor Calibration Uncertainty
1
Frequency band
N1923A
N1924A
50 MHz to 500 MHz
4.5%
4.3%
500 MHz to 1 GHz
4.0%
4.2%
1GHz to 10 GHz
4.0%
4.4%
10 GHz to 18 GHz
5.0%
4.7%
18 GHz to 26.5 GHz
5.9%
26.5 GHz to 40 GHz
6.0%
Physical characteristics
Dimensions
N1923A
N1924A
135 mm x 40 mm x 27 mm (5.3 in x 1.6 in x 1.1 in)
127 mm x 40 mm x 27 mm (5.0 in x 1.6 in x 1.1 in)
Weights with cable
Option 105
Option 106
0.4 kg (0.88 Ib)
0.6 kg (1.32 Ib)
Fixed sensor cable
lengths
Option 105
Option 106
1.5 m (5-feet)
3.0 m (10-feet)
1. Beyond 70 % humidity, an additional 0.6% should be added to these values.
Environmental conditions
General
Complies with the requirements of the EMC Directive 89/336/EEC
Operating
Temperature
0° C to 55° C
Maximum humidity
95% at 40° C (non-condensing)
Minimum humidity
15% at 40 °C (non-condensing)
Maximum altitude
3,000 meters (9,840 feet)
Storage
Non-operating storage temperature –30° C to +70° C
Non-operating maximum humidity
90% at 65 °C (non-condescending)
Non-operating maximum altitude
15,420 meters (50,000 feet)