beautypg.com

8990b peak power analyzer specifications – Atec Agilent-8990B User Manual

Page 12

background image

12

N6904A/8990B-1FP multi-pulse analysis software option

Multi-pulse specifications
Maximum capture frame

512 (each channel 1 and channel 4)

Minimum pulse to pulse duration

1 µs

Number of histogram bins

20 (user adjustable)

Pulse to Pulse measurement

Compare any two pulses from the captured frames

Sensor compatibility

N1921A

P-Series wideband power sensor, 50 MHz to 18 GHz

N1922A

P-Series wideband power sensor, 50 MHz to 40 GHz

N1923A

Wideband power sensor, 50 MHz to 18 GHz

N1924A

Wideband power sensor, 50 MHz to 40 GHz

Computer system and peripherals, I/O ports

Display
Display

15 inch color XGA TFT-LCD with touchscreen capability

Computer system and peripherals
Operating system

Windows 7 Embedded Standard

CPU

Intel

®

Core 2TM Duo CPU E8400 3 GHz microprocessor

System memory

4 GB

Drives

• ≥ 250 GB internal hard disk (option 800)
• ≥ 250 GB removable hard disk (option 801)

Peripherals

• Optical USB mouse and compact keyboard supplied.
• Supports any Windows compatible input device with a PS/2 or USB interface.

File types
Waveforms

Comma separated values (*.csv)

Images

BMP, TIFF, GIF, PNG or JPEG

I/O ports
LAN

RJ-45 connector, supports 10Base-T, 100Base-T and 1000Base-T. Enables web-
enabled remote control, e-mail on trigger, data/file transfers and network printing.

RS-232 (serial)

COM1, printer and pointing device support

PS/2

Two ports. Supports PS/2 pointing and input devices

USB 2.0 Hi-Speed

• Three ports (front panel)

• Four ports (side panel)

• Allows connection of USB peripherals like storage devices and pinitng devices

while the peak power analyzer is turned on. One device port on the side.

Dual-monitor video output

15 pin XGA on side, full color output of scope waveform display or dual monitor
video output

Auxiliary output

DC (± 2.4 V); square wave ~755 Hz with ~200 ps rise time

Trigger out

Output provides TTL compatible logic levels and uses a BNC connector

Time base reference output

• 10 MHz, amplitude into 50 Ω, 800 m Vpp to 12.6 Vpp (4 dBm ±2 dB) if derived from
internal reference.
• Tracks external reference input amplitude ±1 dB if applied and selected.

Time base reference input

10 MHz, input Z = 50 Ω.

Minimum, –2 dBm

Maximum, +10 dBm

Remote programming
Interface

LAN and USB 2.0 interface

Command language

SCPI

8990B Peak Power Analyzer Specifications

(continued)