beautypg.com

General data (all models) – Atec EM-Test-Netwave-3-Phase Series User Manual

Page 3

background image

DATA SHEET > NetWave Series 3-phase > 20130606

TECHNICAL DETAILS

GENERAL DATA (ALL MODELS)

SPECIFICATIONS

Output frequency

DC - 5,000Hz

Frequency
accuracy, stability

100ppm

Output connectors

Safety lab connectors
CEE type 32A (only for NetWave 20.x
and NetWave 30.x)

Interfaces

GPIB
Ethernet
RS 232 (input from DPA analyser)
Frame bus (internal system bus)

REGULATION

Voltage sense

Internal or external, 4 wires

Distortion (THD)

Less than 0.5%, @50/60Hz

Output voltage
Stability

Better than 0.1%

Output voltage
Accuracy

Better than 0.5%

Max.
compensatable
drop on wires

5% f.s.

Current limiter

5A to Imax

Protection

Over current
Over voltage
Over temperature
Low voltage

WAVEFORM GENERATOR

Segment types DC

DC, Ramp, Square, Triangle,
Sawtooth, Step, Sine, Sine sweep,
Sine ramp, Damped sinewave, Sine
ripple, Profile, Square sweep, Noise,
Sine Dwell, Sinc, Harmonic,
Exponent ...

Segment types AC

Sine, Modulation, Sine sweep,
Sweep on Sine, Sine up/down, Sine
unbalance, Overswing, Sine offset,
Sine Dip, Harmonic, Interharmonic,
Interharmonic step, Harmonic
distortion ...

Segment duration

Unlimited

TRIGGER AND DUT MONITORING

Trigger

2 inputs, 2 outputs

DUT monitors

2 inputs, configurable

GENERAL DATA (ALL MODELS)

DISPLAY AND CONTROLS

Display

2-Line LCD, 40 characters

LED indicators

Power On
Active output channel
Trigger
Functional status hard disk

Operation

6 function keys,
Test On key: ON/OFF key for the
power source

DIMENSIONS (ROLLS AND CRANE SUPPORT INCLUDED)

NetWave 20.x

approx. 1785 x 930 x 755mm
approx. 1785 x 1210 x 755mm
(recovery)

NetWave 30.x

approx. 1785 x 930 x 755mm
approx. 1785 x 1210 x 755mm
(recovery)

NetWave 60.x

approx. 2080 x 1205 x 970mm
approx. 2080 x 1615 x 970mm
(recovery)

WEIGHT (ROLLS AND CRANE SUPPORT INCLUDED)

NetWave 20.x

approx. 740kg
approx. 810kg (recovery)

NetWave 30.x

approx. 740kg
approx. 810kg (recovery)

NetWave 60.x

approx. 1,180kg
approx. 1,380kg (recovery)

MAINS

Supply voltage

3 x 400V (3P,N,PE);
optional 3 x 208V (3P,N,PE)

Input current

50A/90A (NetWave 20.x)*
70A/140A (NetWave 30.x)*
140A/212A (NetWave 60.x)*
* the higher figure represents the 3s
short-term current

Line frequency

45Hz - 65Hz

Connectors

Screwed terminals

AMBIENT CONDITIONS

Temperature

0°C - 40°C

Rel. humidity

10% - 90%, non condensing

www.emtest.com

© EM TEST > PAGE 3/6