Performance characteristics, Continued) – Atec Agilent-7000A Series User Manual
Page 21
Performance characteristics
(continued)
Trigger system (continued)
Nth edge burst
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge
count: 65,536.
Autoscale
Finds and displays all active scope and logic (for 7000 Series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
Scope channel triggering
Range (internal)
±6 div from center screen
Sensitivity*
<10 mV/div: greater of 1 div or 5 mV; ≥10 mV/div: 0.6 div
Coupling
AC, ~10 Hz on MSO/DSO701xA/703xA/705xA/710xA,
DC, noise reject, HF reject and LF reject (~50 kHz)
Digital (D15 - D0) channel triggering (7000A Series MSO or MSO-upgraded 7000A Series DSO)
Threshold range (user defined)
±8.0 V in 10 mV increments
Threshold accuracy
±(100 mV + 3% of threshold setting)
Predefined thresholds
TTL = 1.4 V, CMOS = 2.5 V, ECL = –1.3 V
External (EXT) triggering
MSO/DSO7xx2A (2-/2+16-ch models)
MSO/DSO7xx4A (4-/4+16-ch models)
Input impedance
1 MΩ ± 3% || 14 pF or 50 Ω
2.14 kΩ ±5%
Maximum input
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
±15 V
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-Ω input
Range
DC coupling: trigger level ±1 V and ±8 V
±5 V
Sensitivity
For ±1 V range setting:
DC to 500 MHz: 500 mV
DC to 100 MHz, 100 mV;
For ±8 V range setting:
DC to 100 MHz, 250 mV;
>100 MHz to bandwidth of oscilloscope: 500 mV
Coupling
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
Probe ID
Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
21