beautypg.com

Performance characteristics – Atec Agilent-7000A Series User Manual

Page 16

background image

Performance characteristics

Acquisition: scope channels

Sample rate

MSO/DSO701xA: 2 GSa/s each channel

MSO/DSO703xA: 2 GSa/sec each channel

MSO/DSO705xA/710xA: 4 GSa/sec half channel*, 2 GSa/sec each channel

Equivalent-time sample rate: 400 GSa/s (when real-time mode is turned off)

Memory depth

2 channels/4 channels

Standard

8 Mpts/4 Mpts

Vertical resolution

8 bits

Peak detection

MSO/DSO701xA: 500-ps peak detect

MSO/DSO703xA: 500-ps peak detect

MSO/DSO705xA/710xA: 250-ps peak detect

Averaging

Selectable from 2, 4, 8, 16, 32, 64 … to 65536

High resolution mode

Average mode with avg = 1

12 bits of resolution when ≥10 µs/div at 4 GSa/s or ≥20-µs/div at 2 GSa/s

Filter

Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope,

whichever is less) with vectors on and in real-time mode

Acquisition: digital channels (7000A Series MSO or MSO-upgraded 7000A Series DSO)

Sample rate

2 GSa/sec one pod**, 1 GSa/sec each pod

Maximum input frequency

250 MHz

Memory depth

One pod/both pods (with scope channels turned off)

Standard

8 Mpts/4 Mpts

One pod/both pods (with scope channels turned on)

Standard

2.5 Mpts/ 1.25 Mpts

Vertical resolution

1 bit

Glitch detection

2 ns (min pulse width)

* Half channel is when one of channel 1 or 2 is turned on, and/or one of channel 3 or 4 is turned on.

** A pod is a group of eight digital channels, either 0-7 or 8-15.

16