Performance characteristics

(continued)

Horizontal

Range

MSO/DSO701xA: 2 nsec/div to 50 sec/div

MSO/DSO703xA: 2 nsec/div to 50 sec/div

MSO/DSO705xA: 1 nsec/div to 50 sec/div

MSO/DSO710xA: 500 psec/div to 50 sec/div

Resolution

2.5 ps

Time scale accuracy*

≤ ± (15+2* (instrument age in years)) ppm

Vernier

1-2-5 increments when off, ~25 minor increments between major settings when on

Delay range

Pre-trigger (negative delay): Greater of 1 screen width or 1 ms

Post-trigger (positive delay): 1 s to 500 seconds

Analog delta-t accuracy

Same channel: ±0.0015% reading ±0.1% screen width ±20 ps

Channel-to-channel: ±0.0015% reading ±0.1% screen width ±40 ps

Same channel example (MSO/DSO705xA):

For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),

delta-t accuracy = ±{0.0015% (10 µs) + 0.1% (50 µs) + 20 ps} = 50.17 ns

Logic delta-t accuracy

Same channel: ±0.005% reading ±0.1% screen width ±(1 logic sample period, 1 ns)

Channel-to-channel:

±0.005% reading ±0.1% screen width ±(1 logic sample period) ±chan-to-chan skew

Same channel example:

For signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),

delta-t accuracy = ±{0.005% (10 µs) + 0.1% (50 µs) + 1 ns} = 51.5 ns

Modes

Main, zoom, roll, XY, segmented (optional)

XY

Bandwidth: Max bandwidth

Phase error at 1 MHz: < 0.5 degrees

Z Blanking: 1.4 V blanks trace (use external trigger on MSO/DSO7xx2A,

channel 4 on MSO/DSO7xx4A)

Reference positions

Left, center, right

Segmented memory re-arm time

8 µs (minimum time between trigger events)

Trigger system

Sources

MSO7xx2A: Ch 1, 2, line, ext, D15 - D0

DSO7xx2A: Ch 1, 2, line, ext

MSO7xx4A: Ch 1, 2, 3, 4, line, ext, D15 - D0

DSO7xx4A: Ch 1, 2, 3, 4, line, ext

Modes

Auto, normal (triggered), single

Holdoff time

~60 ns to 10 seconds

Trigger jitter

15 ps rms

* Denotes warranted specification. Specifications are valid after a 30 minute warm-up period and ±10 °C from firmware calibration procedure.

19