Infiniium 9000 series performance characteristics – Atec Agilent-Infiniium-9000 Series User Manual
Page 20
20
Infiniium 9000 Series performance characteristics
Acquisition
9104, 9254, 9404
9064
Maximum real-time sample rate
4 ch x 10 GS/s or 2 ch x 20 GS/s
4 ch x 5 GSa/s or 2 ch x 10 GSa/s
Memory depth per channel
Standard
20 Mpts on 4 channels, 40 Mpts on 2 channels
Option 50M
50 Mpts on 4 channels, 100 Mpts on 2 channels
Option 100
100 Mpts on 4 channels, 200 Mpts on 2 channels
Option 200
200 Mpts on 4 channels, 400 Mpts on 2 channels
Option 500
500 Mpts/ 250 Mpts on 4 channels, 1 Gpts/ 500 Mpts on 2 channels (single/repetitive mode)
Sampling Modes
Real-time
Real-time with peak detect
Real-time with high resolution (user selectable to 9-, 10-, 11-, or 12-bits of resolution)
Real-time with roll mode (200 ms to 20 sec.)
Equivalent-time (1.0 ps fine interpolator resolution yields a maximum effective sample rate of 1,000 GSa/s)
Segmented memory ( 1 ps time stamp resolution between segments)
Up to 8192 segments for 20 Mpts standard memory, up to 131,072 segments with Option 500
Maximum time between triggers is 562,950 seconds (6.5 days)
Re-arm time (minimum time between trigger events) is 4.5 µs with analog channels, 5.8 µs with digital channels on
Filters
Sin (x) / x Interpolation
Acquisition: digital channels
Maximum real time sample rate
2 GSa/s
Maximum memory depth per channel
128/ 64 Mpts with 2 GSa/s. 64/32 Mpts with sampling < 2 GSa/s (single/repetitive mode).
Minimum width glitch detection
2 ns
Trigger: scope channels
Trigger sources
Channel 1, channel 2, channel 3, channel 4, aux, and line
Sensitivity
1 MΩ input, edge trigger,
DC to 500 MHz: 0.6 div
50 Ω
DC to 2 GHz, 0.5 div
2 GHz to 4 GHz: 1.0 div
Auxiliary
DC to 700 MHz: 300 mVp-p
Trigger level range
± 4 div from center screen (50 Ω)
Channel 1,2,3,4
± 8 div from center screen with max of ± 8 V (1 MΩ)
Auxiliary
± 5 V (50Ω up to 500 MHz with at least 500 mV signal swing)
Sweep modes
Auto, triggered, single
Display jitter (displayed trigger jitter)
1,2
Trigger holdoff range
100 ns to 10 s fixed and random
Trigger actions
Specify an action to occur, and the frequency of the action, when a trigger condition occurs.
Actions include: e-mail on trigger and execute “multipurpose” user settings
Trigger coupling 1 MΩ: DC, AC, (10 Hz) low frequency reject (50 kHZ high pass filter), high frequency reject (50 kHz low pass filter)
1. Internal edge trigger mode. Trigger threshold = fixed voltage at 50% level. The slew rate independent value in the formula represents the traditional trigger jitter.
2. Display jitter example. At 100 mV/div typical noise values are 3.2 mV RMS for 9404 models, 2.5 mV RMS for 9254A models, and 2.4 mV RMS for 9104A models. For slew rate of 500 mVpp
sin wave with frequency equal to max analog bandwidth of each model, typical display jitter is .95 ps RMS for 9404A models, .97ps for 9254A models, and 1.7 ps RMS for 9104A models.
9064 .50
9104 .35
9254 .50
9404 .40