beautypg.com

Infiniium 9000 series performance characteristics – Atec Agilent-Infiniium-9000 Series User Manual

Page 17

background image

17

Infiniium 9000 Series performance characteristics

Vertical: scope channels

9064A

9104A

9254A

9404A

Analog bandwidth (–3 dB) 50 Ω

1

600 MHz

1.0 GHz 2.5 GHz 4 GHz

1MΩ

500 MHz

500 MHz

500 MHz

500 MHz

Typical Rise Time / Fall

540 ps

253 ps

142 ps

85 ps

Time 10% to 90% at 50 Ω

Typical Rise Time / Fall

360 ps

174 ps

98 ps

59 ps

Time 20% to 80% at 50 Ω

Input channels

DSO9000 – 4 analog

MSO9000 – 4 analog + 16 digital

Input impedance

1

50 Ω ± 2.5%, 1 MΩ ± 1% (11pF typical)

Input sensitivity

3

1 MΩ: 1 mV/div to 5 V/div

50 Ω: 1 mV/div to 1 V/div

Input coupling

1 MΩ: AC (3.5 Hz), DC

50 Ω:DC

Bandwidth limit

20 MHz on 1 MΩ input ; 500 MHz up to full scope bandwidth in increments of 500 MHz

Vertical resolution

2,3

8 bits, ≥12 bits with averaging

Channel-to-channel isolation

DC to 50 MHz: 50 dB

>50 MHz to 2.5 GHz: 40 dB

>2.5 GHz to 4 GHz: 25 dB

DC gain accuracy

1,2,3

± 2% of full scale at full resolution on channel scale ± 5 ˚C from cal temp

Maximum input voltage

1

1 MΩ: 150V RMS or DC, CAT I

± 250 V (DC + AC) in AC coupling

50 Ω: 5 Vrms

Offset range

Vertical sensitivity

Available offset

1 MΩ

1 mV to <10 mV/div

± 2 V

10 mV to <20 mV/div

± 5 V

20 mV to <100 mV/div

± 10 V

100 mV to <1 V/div

± 20 V

1 V to 5 V/div

± 100 V

50 Ω

±12 div or ±4V, whichever is smallest

1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±5 °C from firmware calibration temperature. Input

impedance is valid when V/div scaling is adjusted to show all waveform vertical values within scope display.

2 Vertical resolution for 8 bits = 0.4% of full scale, for 12 bits = 0.024% of full scale.
3 50Ω input: Full scale is defined as 8 vertical divisions.  Magnification is used below 10mV/div, full-scale is defined as 80 mV.  The major scale settings are 5mV, 10mV, 20mV, 50

mV, 100mV, 200 mV, 500 mV, 1V.

1MΩ input: Full scale is defined as 8 vertical divisions.  Magnification is used below 5mV/div, full-scale is defined as 40 mV.  The major scale settings are 5mV, 10mV, 20mV, 50

mV, 100 mV, 200 mV, 500 mV, 1V,2V, 5V.