beautypg.com

Infiniium 9000 series performance characteristics – Atec Agilent-Infiniium-9000 Series User Manual

Page 19

background image

19

Horizontal

Channel-to-channel skew (digital)

2 ns typical

Glitch detect (digital)

≥ 2.0 ns

Main time base range

5 ps/div to 20 s/div

Horizontal position range

0 to ± 200 s

Delayed sweep range

1 ps/div to current main time base setting

Resolution

1 ps

Modes

Main, delayed, roll (200 ms to 20 sec)

Reference positions

Left, center, right

Channel deskew

-1 ms to +1 ms range

Time scale accuracy (internal reference)

Horizontal time base setting ± ((Horizontal time base setting) *(0.4 + 0.5* years since

(External reference clock = off)

calibration)) ppm

Infiniium 9000 Series performance characteristics

Delta-time measurement accuracy

2,3,4,5

1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±5 °C from firmware calibration temperature.
2 Noise is the displayed noise floor. SlewRate is the displayed slew rate of the signal at the threshold crossings. Sample rate = max, sin(x)/x interpolation enabled.
3 Measurement threshold = fixed voltage at 50% level.
4 Time ranges ≤ 10 μs.
5 Values represent time error between two edges on a single channel. Standard deviation value refers to the standard deviation of 256 consecutive measurements performed
using an individual instrument. Reading is the displayed DTMA measurement value.

9064 4.8 20
9104 4.8 15
9254 4.0 15
9404 5.0 20

9064 .33 .1
9104 .33 .05
9254 .33 .10
9404 .35 .15

9064 .75
9104 .65
9254 .75
9404 .80

9064 1.0 1.0
9104 1.0 0.5
9254 .95 1.1
9404 .95 1.2

9064 .75
9104 .65
9254 .75
9404 .80

9064 1.8
9104 1.4
9254 1.9
9404 2.0

Absolute averaging disabled

Absolute >256 averages

Standard deviation averaging disabled

Standard deviation >256 averages

Jitter measurements floor

2,3

Time interval error

4

N-cycle, cycle-cycle jitter

Period jitter