Infiniium 9000 series performance characteristics – Atec Agilent-Infiniium-9000 Series User Manual
Page 19
19
Horizontal
Channel-to-channel skew (digital)
2 ns typical
Glitch detect (digital)
≥ 2.0 ns
Main time base range
5 ps/div to 20 s/div
Horizontal position range
0 to ± 200 s
Delayed sweep range
1 ps/div to current main time base setting
Resolution
1 ps
Modes
Main, delayed, roll (200 ms to 20 sec)
Reference positions
Left, center, right
Channel deskew
-1 ms to +1 ms range
Time scale accuracy (internal reference)
Horizontal time base setting ± ((Horizontal time base setting) *(0.4 + 0.5* years since
(External reference clock = off)
calibration)) ppm
Infiniium 9000 Series performance characteristics
Delta-time measurement accuracy
2,3,4,5
1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±5 °C from firmware calibration temperature.
2 Noise is the displayed noise floor. SlewRate is the displayed slew rate of the signal at the threshold crossings. Sample rate = max, sin(x)/x interpolation enabled.
3 Measurement threshold = fixed voltage at 50% level.
4 Time ranges ≤ 10 μs.
5 Values represent time error between two edges on a single channel. Standard deviation value refers to the standard deviation of 256 consecutive measurements performed
using an individual instrument. Reading is the displayed DTMA measurement value.
9064 4.8 20
9104 4.8 15
9254 4.0 15
9404 5.0 20
9064 .33 .1
9104 .33 .05
9254 .33 .10
9404 .35 .15
9064 .75
9104 .65
9254 .75
9404 .80
9064 1.0 1.0
9104 1.0 0.5
9254 .95 1.1
9404 .95 1.2
9064 .75
9104 .65
9254 .75
9404 .80
9064 1.8
9104 1.4
9254 1.9
9404 2.0
Absolute averaging disabled
Absolute >256 averages
Standard deviation averaging disabled
Standard deviation >256 averages
Jitter measurements floor
2,3
Time interval error
4
N-cycle, cycle-cycle jitter
Period jitter