beautypg.com

T wo/three zone thermal shock, Hot/cold or hot/ambient/cold tests possible – Atec Espec-ESX Series User Manual

Page 16

background image

Two/Three Zone
Thermal Shock

Hot/cold or hot/ambient/cold
tests possible

Small footprint and large workspace

Traditional three zone thermal shock chambers can

be quite large, as three separate chambers are used.

The TSA series has just one chamber, circulating

cold, hot, or ambient air, as needed.

The conditioned air can be pre-cooled or pre-heated

beyond the setpoint, allowing quicker recovery of

the desired condition after transfer.

Test samples don’t move.

Horizontal airflow design assures uniform results

Extremely compact systems with a vertical

sliding door for added space savings

The touch-screen controller is specifically

designed for thermal shock testing

TSA-71S-W (4.3'w x 4.5'd x 6.3'h exterior)

T

O

U

C

H

SC

REEN

P R O G RA

M

M

E

R

• Catalog available

Performance example

Temperature

Test Load

Model

Range

Temperature Range

Recovery Time

IC Chips

Interior Volume

TSA-71H-W

Two zone

70 L

-65 to 150

°

C

(2.5 cu. ft.)

Mil-Std 883 1010.7 C

TSA-71S-A

TSA-71S-W

TSA-101S-W

TSA-201S-W

TSA-41L-A

TSA-71L-A

TSA-101L-A

TSA-301L-W

Hot exposure:

60 to 200

°

C

(140 to 392

°

F)

Ambient

exposure

Cold exposure

-75 to 0

°

C

(-103 to 32

°

F)

Hot exposure:

60 to 200

°

C

(140 to 392

°

F)

Ambient

exposure

Cold exposure:

-65 to 0

°

C

(-85 to 32

°

F)

Three zone

-65 to ambient

to 150

°

C

Three zone

-55 to ambient

to 150

°

C

15 min. worst-case

10 kg (22 lb.)

product recovery

5 min.

upstream air

10 min. upstream air

5 min. upstream air

10 min. upstream air

20 kg (44 lb.)

24 kg (53 lb.)

5 kg (11 lb.)

20 kg (44 lb.)

2.5 kg (5.5 lb.)

70 L

(2.5 cu. ft.)

200 L

(7.1 cu. ft.)

40 L

(1.4 cu. ft.)

70 L

(2.5 cu. ft.)

110 L

(3.9 cu. ft.)

300 L

(10.6 cu. ft.)

110 L

(3.9 cu. ft.)

15

Go ESPEC: www.espec.com