beautypg.com

Bluetooth edr receiver tests, Edr rx sensitivity, Edr rx ber floor sensitivity – Atec Agilent-N4010A User Manual

Page 6: Edr rx maximum input level

background image

6

Bluetooth EDR receiver tests

EDR Rx sensitivity

Link conditions

Link mode

Test mode (loopback)

Payload

PRBS9, BSOO, BSFF, BS55

Packet type

2-DH1, 2-DH3, 2-DH5, 3-DH1,

3-DH3, 3-DH5

Number of bits

1 to 200,000,000

Impairments

Frequency offset

±100 kHz

Frequency offset resolution 1 kHz

Symbol timing

–30 to +30 ppm

Symbol timing resolution

1 ppm

“Dirty transmitter” impairments for EDR Rx sensitivity

measurements

Set of

parameters

Carrier offset

frequency (kHz)

Symbol timing

offset (ppm)

1

0

0

2

+65

+20

3

–65

–20

Sine impairments for EDR Rx sensitivity measurements

Modulation frequency range 300 Hz to 10 kHz

Resolution

100 Hz

Maximum deviation range

0 Hz to 40 kHz

Resolution

1 kHz

Measurement

Supported measurements

BER, number of bit errors, number

of Rx bits, PER, number of NACK

packets, number of errored packets,

number of Tx packets, number of

HEC, CRC, and NACK error

Number of measurement

channels

1

3, hopping

Range

0 to –90 dBm

Resolution

0.1 dB

Accuracy

2, 3

±0.6 dB, –35 to –90 dBm

±1 dB, > 35 dBm

EDR Rx BER floor sensitivity

Link conditions

Link mode

Test mode (loopback)

Hopping

On or off

Payload

PRBS9, BSOO, BSFF, BS55

Packet type

2-DH1, 2-DH3, 2-DH5, 3-DH1,

3-DH3, 3-DH5

Number of bits

1 to 200,000,000

Measurement

Supported measurements

BER, number of bit errors, number

of Rx bits, PER, number of NACK

packets, number of errored packets,

number of Tx packets, number of

HEC, CRC, and NACK error

Number of measurement

channels

1

3, hopping

Range

0 to –90 dBm

Resolution

0.1 dB

Accuracy

2, 3

±0.6 dB, –35 to –90 dBm

±1 dB, > 35 dBm

EDR Rx maximum input level

Link conditions

Link mode

Test mode (loopback)

Hopping

On or off

Payload

PRBS9, BSOO, BSFF, BS55

Packet type

2-DH1, 2-DH3, 2-DH5, 3-DH1,

3-DH3, 3-DH5

Number of bits

1 to 200,000,000

Measurement

Supported measurements

BER, number of bit errors, number

of Rx bits, PER, number of NACK

packets, number of errored packets,

number of Tx packets, number of

HEC, CRC, and NACK error

Number of measurement

channels

1

3, hopping

Range

0 to –90 dBm

Resolution

0.1 dB

Accuracy

2, 3

±0.6 dB, –35 to –90 dBm

±1 dB, > 35 dBm

1. Internal sequencer enables three measurement channels to be measured

consecutively. Measurements on all 79 Bluetooth channels are supported.

2. Verified using CW measurements.
3. Add 0.01 dB/°C from 30 to 55 °C, add 0.07 dB/°C from 20 to 0 °C.