Bluetooth Specifications

N4010A Option 101 Bluetooth

• Provides ability to act as a Bluetooth master, perform

inquiry, and establish a connection in test mode or

normal mode
• Makes measurements in accordance with Bluetooth RF

Test Specification 1.2
• Integral sequencer allows test plans to be created and

edited easily
• All tests default to SIG standard settings – user may

change settings to match particular test requirements

Bluetooth tests

1

Output power

Link conditions

Link mode

Test mode (loopback, Tx), normal

mode (ACL, SCO)

Hopping

2

On or off

Packet type

2

DH1, DH3, DH5, HV3

Payload

2

PRBS9, BSOO, BSFF, BSOF, BS55

Measurement

Supported measurements

Average power, peak power

Number of measurement

channels

3

3

Range

+23 to –70 dBm

Measurement resolution

0.01 dB

Measurement accuracy

±0.5 dB

Power control

Link conditions

Link mode

Test mode (loopback, Tx)

Hopping

On or off

Packet type

DH1, DH3, DH5, HV3

Payload

PRBS9, BSOO, BSFF, BSOF, BS55

Measurement

Supported measurements

Average power, min/max step size

Number of measurement

channels

3

3

Range

+23 to –70 dBm

Measurement resolution

0.01 dB

Measurement accuracy

±0.5 dB

Modulation characteristics

Link conditions

Link mode

Test mode (loopback, Tx), normal

mode (ACL, SCO)

Hopping

2

On or off

Packet type

2

DH1, DH3, DH5, HV3

Payload

2

BS55, BSOF

Measurement

Supported measurements

Min/max ∆f1

avg

, min ∆f2

max

(kHz),

total ∆f2

max

> ∆f2

max

lower limit

(%) min of min ∆f2

avg

/ max ∆f1

avg

,

pseudo frequency deviation (∆f1

and ∆f2) in normal mode

Number of measurement

channels

3

3

RF input level range

+23 to –70 dBm

Deviation range

–400 to +400 kHz

Deviation resolution

100 Hz

Ratio resolution

0.1%

Measurement accuracy

4

As frequency reference ±100 Hz

1. Performance of the N4010A signal source or signal analyzer over

wider temperature (specified later in this document) applies to all

Bluetooth tests listed.

2. Normal mode measurements made with hopping on, NULL packet,

and no payload.

3. Internal sequencer enables three measurement channels to be

measured consecutively. Measurements on all 79 Bluetooth channels

are supported.

4. Example, using the 10 MHz reference with accuracy of 10 Hz (1 ppm),

at frequency of 2.402 GHz, frequency accuracy would be in the range

of ±((2.402 GHz x 10 Hz)/10 MHz) ± 100 Hz = ±2402 Hz ± 100 Hz =

±2502 Hz.