Pc hardware specifications, N4010a general specifications, Inputs/outputs – Atec Agilent-N4010A User Manual
Page 13: Size and weight, Regulatory information
![background image](/manuals/317902/13/background.png)
13
PC Hardware specifications
• Microsoft
®
Windows
®
2000 and XP
®
only
• 2.4 GHz Pentium
®
or equivalent minimum, 2.8 GHz
recommended
• 200 MH available on hard drive
• 256 MB RAM minimum, 500 MB RAM recommended
• USB 2.0, TCP-IP LAN, or GPIB connection to test set
• Agilent I/O Libraries Suite 14.1 or greater. For
information on Agilent I/O Libraries Suite features and
installation requirements, please go to:
Frequency reference
Frequency
10 MHz
Accuracy
20 to 30 °C
0 to 55 °C
±1 x 10
-6
(±1 ppm)
±1.5 x 10
-6
(±1.5 ppm)
Aging (first year)
±1 x 10
-6
/year
10 MHz input
BNC(f), 50 W
10 MHz output
BNC(f), 50 W
Power requirements
Voltage
100 to 240 VAC, 47 to 63 Hz
Power
150 VA maximum
Environmental
Operating temperature
0 to 55 °C
Storage temperature
–40 to +70 °C
Operating humidity
15 to 95% relative humidity
(non-condensing)
EMI compatibility
Radiated emission is in compliance
with CISPR Pub 11/1990 Group 1
Class A
Inputs/outputs
Front panel
RF input/output
Type-N (f), 50 W
Rear panel
10 MHz REF IN
BNC(f), 50 W
10 MHz REF OUT
BNC(f), 50 W
GPIB
IEEE-488
LAN
RJ-45, 10/100-T
USB
USB 1.0/2.0
Additional rear panel connectivity with N4010A input/output
connectivity Option 110
AUX RF input/output
Type-N (f), 50 W
TRIG IN
BNC(f), 50 W; input has TTL
compatible logic levels
TRIG OUT
BNC(f), 50 W; output has TTL
compatible logic levels
75 MHz IF output
SMA (f), 50 W
Event 1
BNC(f), 50 W
Event 2
BNC(f), 50 W
Bluetooth and WLAN triggers,
data, and clock
25-way D (f)
Size and weight
Dimensions
(H x W x D)
With handle and bumpers
105 mm x 370 mm x 390 mm
Without handles and
bumpers
105 mm x 330 mm x 375 mm
Weight
5.9 kg (12.98 lbs) for N4010A-101
7.2 kg (15.84 lbs) for N4010A-102,
103
Regulatory information
Product safety
Conforms to the following product
specifications:
IEC61010-1:2001/
EN61010-1:2001
CAN/CSA-C22.2 No 1010.1-92
Low voltage directive 72/23/EEC
General conditions
The conformity assessment
requirements have been met
using the technical construction
file route for compliance with the
requirements of the EMC Directive
89/336/EEC