beautypg.com

Initial carrier frequency tolerance, Carrier frequency drift, Sensitivity – single slot packets – Atec Agilent-N4010A User Manual

Page 3

background image

3

Initial carrier frequency tolerance

Link conditions

Link mode

Test mode (loopback, Tx), normal

mode (ACL)

Hopping

1

On or off

Packet type

1

DH1, DH3, DH5, HV3

Payload

1

PRBS9, BSOO, BSFF, BSOF, BS55

Measurement

Supported measurements

Maximum and minimum error/

channel

Number of measurement

channels

2

3

RF input level range

+23 to –70 dBm

Frequency

Nominal channel freq ±150 kHz

Measurement accuracy

3

As frequency reference ±100 Hz

Carrier frequency drift

Link conditions

Link mode

Test mode (loopback, Tx), normal

mode (ACL)

Hopping

1

On or off

Packet type

1

DH1, DH3, DH5, HV3

Payload

1

PRBS9, BSOO, BSFF, BSOF, BS55

Measurement

Supported measurements

Maximum and minimum measure-

ments drift at each frequency

during the test, pseudo frequency

drift in normal mode

Number of measurement

channels

2

3

RF input level range

+23 to –70 dBm

Measurement range

±100 kHz

Measurement accuracy

3

As frequency reference ±100 Hz

Sensitivity – single slot packets

Link conditions

Link mode

Test mode (loopback, Tx), normal

mode (ACL)

Hopping

1

On or off

Packet type

1

DH1, DH3, DH5

Payload

1

PRBS9, BSOO, BSFF, BSOF, BS55

Number of bits

1 to 200,000,000

Impairments – default to table

Frequency offset

±75 kHz

Modulation index

0.28 to 0.35

Modulation index resolution 0.01

Symbol timing

–20 ppm, 0, +20 ppm

Symbol timing resolution

1 ppm

Measurement

Supported measurements

BER, number of bit errors, number

of Rx bits, PER, number of NACK

packets, number of errored packets,

number of Tx packets, PER only in

normal mode

Number of measurement

channels

2

3, hopping

Range

0 to –90 dBm

Resolution

0.1 dB

Accuracy

4, 5

±0.6 dB, –35 to –90 dBm,

±1 dB, > 35 dBm

Sine impairments (applicable for single slot packets,

multi-slot packets, and maximum input level)

Modulation frequency range 300 Hz to 1.6 kHz

Resolution

100 Hz

Maximum deviation range

0 Hz to 40 kHz

Resolution

1 kHz

‘Dirty transmitter’ impairments table for Rx sensitivity

tests (applicable for single slot packets, multi-slot

packets, and maximum input level)

Set of

parameters

Carrier

frequency

offset (kHz)

Modulation

index

Symbol timing

error (ppm)

1

75

0.28

–20

2

14

0.30

–20

3

–2

0.29

+20

4

1

0.32

+20

5

39

0.33

+20

6

0

0.34

–20

7

–42

0.29

–20

8

74

0.31

–20

9

–19

0.28

–20

10

–75

0.35

+20

1. Normal mode measurements made with hopping on, NULL packet, and

no payload.

2. Internal sequencer enables three measurement channels to be measured

consecutively. Measurements on all 70 Bluetooth channels are supported.

3. Example, using the 10 MHz reference with accuracy of 10 Hz (1 ppm),

at frequency of 2.402 GHz, frequency accuracy would be in the range

±((2.402 GHz x 10 Hz)/10 MHz) ± 100 Hz = ±2402 Hz ± 100 Hz = ±2502 Hz.

4. Verified using CW measurements.
5. Add 0.01 dB/°C from 30 to 55 °C, add 0.07 dB/°C from 20 to 0 °C.