beautypg.com

Cla-Val 49-01/649-01 Technical Manual User Manual

Page 9

background image

Diaphragm Check (#1 )

1. Shut off pressure to the Hytrol Valve by slowly closing upstream

and downstream isolation valves. SEE CAUTION.
2. Disconnect or close all pilot control lines to the valve cover and

leave only one fitting in highest point of cover open to atmosphere.
3.With the cover vented to atmosphere, slowly open upstream

isolation valve to allow some pressure into the Hytrol Valve body.

Observe the open cover tapping for signs of continuous flow. It is

not necessary to fully open isolating valve. Volume in cover cham-

ber capacity chart will be displaced as valve moves to open posi-

tion. Allow sufficient time for diaphragm assembly to shift posi-

tions. If there is no continuous flow, you can be quite certain the

diaphragm is sound and the diaphragm assembly is tight. If the

fluid appears to flow continuously this is a good reason to believe

the diaphragm is either damaged or it is loose on the stem. In

either case, this is sufficient cause to remove the valve cover and

investigate the leakage. (See “Maintenance” Section for procedure.)

Freedom of Movement Check (#2)

4. Determining the Hytrol Valve’s freedom of movement can be

done by one of two methods.
5. For most valves it can be done after completing Diaphragm

Check (Steps 1, 2, and 3). SEE CAUTION. At the end of step 3

the valve should be fully open.
6. If the valve has a Cla-Val X101 Position Indicator, observe the

indicator to see that the valve opens wide. Mark the point of max-

imum opening.
7. Re-connect enough of the control system to permit the appli-

cation of inlet pressure to the cover. Open pilot system cock so

pressure flows from the inlet into the cover.
8. While pressure is building up in the cover, the valve should

close smoothly. There is a hesitation in every Hytrol Valve closure,

which can be mistaken for a mechanical bind. The stem will

appear to stop moving very briefly before going to the closed posi-

tion. This slight pause is caused by the diaphragm flexing at a

particular point in the valve’s travel and is not caused by a

mechanical bind.
9. When closed, a mark should be made on the X101 Valve posi-

tion indicator corresponding to the “closed” position. The distance

between the two marks should be approximately the stem travel

shown in chart.

10. If the stroke is different than that shown in stem travel chart

this is a good reason to believe something is mechanically restrict-

ing the stroke of the valve at one end of its travel. If the flow does

not stop through the valve when in the indicated “closed” position,

the obstruction probably is between the disc and the seat. If the

flow does stop, then the obstruction is more likely in the cover. In

either case, the cover must be removed, and the obstruction locat-

ed and removed. The stem should also be checked for scale build-

up. (See “Maintenance, section for procedure.)
11. For valves 6” and smaller, the Hytrol Valve’s freedom of move-

ment check can also be done after all pressure is removed from

the valve. SEE CAUTION. After closing inlet and outlet isolation

valves and bleeding pressure from the valve, check that the cover

chamber and the body are temporarily vented to atmosphere.

Insert fabricated tool into threaded hole in top of valve stem, and

lift the diaphragm assembly manually. Note any roughness. The

diaphragm assembly should move smoothly throughout entire

valve stroke. The tool is fabricated from rod that is threaded on

one end to fit valve stem and has a “T” bar handle of some kind

on the other end for easy gripping. (See chart in Step 4 of

“Disassembly” Section.)
12. Place marks on this diaphragm assembly lifting tool when the

valve is closed and when manually positioned open. The distance

between the two marks should be approximately the stem travel

shown in stem travel chart. If the stroke is different than that

shown, there is a good reason to believe something is mechani-

cally restricting the stroke of the valve. The cover must be

removed, and the obstruction located and removed. The stem

should also be checked for scale build-up. (See “Maintenance”

Section for procedure.)

Tight Sealing Check (#3)

13. Test for seat leakage after completing checks #1 & #2 (Steps

1 to 12). SEE CAUTION. Close the isolation valve downstream of

the Hytrol Valve. Apply inlet pressure to the cover of the valve, wait

until it closes. Install a pressure gauge between the two closed

valves using one of the two ports in the outlet side of the Hytrol.

Watch the pressure gauge. If the pressure begins to climb, then

either the downstream isolation valve is permitting pressure to

creep back, or the Hytrol is allowing pressure to go through it.

Usually the pressure at the Hytrol inlet will be higher than on the

isolation valve discharge, so if the pressure goes up to the inlet

pressure, you can be sure the Hytrol is leaking. Install another

gauge downstream of isolating valve. If the pressure between the

valves only goes up to the pressure on the isolation valve

discharge, the Hytrol Valve is holding tight, and it was just the iso-

lation valve leaking.

STEM TRAVEL

(Fully Open to Fully Closed)

Valve Size (inches)

Travel (inches)

Inches

MM

Inches

MM

1 1/4

32

0.4

10

1 1/2

40

0.4

10

2

50

0.6

15

2 1/2

65

0.7

18

3

80

0.8

20

4

100

1.1

28

6

150

1.7

43

8

200

2.3

58

10

250

2.8

71

12

300

3.4

86

14

350

4.0

100

16

400

4.5

114

24

600

6.5

165

30

800

7.5

190

36

900

8.5

216

COVER CHAMBER CAPACITY

(Liquid Volume displaced when valve opens)

Valve size (inches)

Displacement

Gallons

Liters

1 1/4

.020

.07

1 1/2

.020

.07

2

.032

.12

2 1/2

.043

.16

3

.080

.30

4

.169

.64

6

.531

2.0

8

1.26

4.8

10

2.51

9.5

12

4.00

15.1

14

6.50

24.6

16

9.57

36.2

24

29.00

109.8

30

42.00

197.0

36

90.00

340.0

3