beautypg.com

Xylem VARIO WATERPROOF PH METER User Manual

Page 45

background image

VARIO pH

What to do if...

45

ba75373e04

07/2006

Display of

l

Instrument does

not react

to keystroke

Sensor symbol

flashes

Message

StoFull

StoFull

StoFull

StoFull

Adapter with electrode
+ measuring solution:

– Conductivity too low

– Use suitable electrode

– Temperature too high

– Use suitable electrode

– Organic liquids

– Use suitable electrode

Cause

Remedy

– Battery almost empty

– Exchange the battery (see

section 5.2)

Cause

Remedy

– Operating condition undefined or

EMC load unallowed

– Processor reset:

Press <

1

> and <

8

> at the

same time and release them
again.
Stored data and calibration
values are retained.

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Clear the memory (see section

4.5.2).