VARIO pH

What to do if...

43

ba75373e04

07/2006

6

What to do if...

Error message

0FL

0FL

0FL

0FL

,

UFL

UFL

UFL

UFL

Error message

Err3

Err3

Err3

Err3

Cause

Remedy

– Air bubble in front of the junction

– Remove air bubble

– Electrode defective

– Replace electrode

Cause

Remedy

Erroneous calibration
The data of the electrode are outside
the specified characteristics.

– Recalibrate

– Reset to default setting (see

section 4.9)

Electrode:

– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

– Moisture in the plug

– Dry plug

– Electrode obsolete

– Replace electrode

– Electrode broken

– Replace electrode

Meter:

– Incorrect buffer set

– Select correct buffer set

– Incorrect solution temperature

(without temperature sensor)

– Set the correct temperature

– Socket damp

– Dry socket

Buffer solutions:

– Incorrect buffer set

– Use a suitable buffer set

– Buffer solutions too old

– Use only once.

Note the shelf life