ba75373e04
07/2006
4.9
Reset
The following measurement parameters are reset to the delivery status:
Note
The calibration data gets lost when the measurement parameters are reset.
Recalibrate after performing a reset. Stored data are retained.
Resetting the
measurement
parameters
Parameter
Default settings
Measuring mode
pH
Asymmetry
0 mV
Slope -59.2
mV/pH
Calibration procedures
TEC
Temperature, manual
25 °C
1
Using <
C_
>, start the INIT initialization menu.
2
Using <
<
> or <
>
> or <
§
>, select YES for initialization.
3
Confirm with <
g
>.
The reset is performed.
The meter switches to the setting of the calibration interval. The
display flashes. You can change the calibration interval (see section
4.4.3) or switch to the measured value display with <
E
>.
0
d
Ident
CLR
Day.Month
MEM
CAL
C
pH
LoBat
min:s Lap
mV
h:min
RESET
LAP
ESC
n i t
S
E
y
I
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- IQ SensorNet ViSolid 700 IQ (SW) (58 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)