12 what to do if, 1 calibration and measuring, Calibration and measuring – Xylem WQL-pH pH DATALOGGERS User Manual
Page 67

WQL-pH
What to do if...
141
ba75821e01
10/2009
12
What to do if...
12.1
Calibration and measuring
Lengthy stability
check with
AutoRead
Implausible
measured values
Logger does not
react
System cannot be
calibrated
Cause
Remedy
No stable measured value
Provide stable measuring conditions
(e.g. temperature)
Cause
Remedy
– Electrode not installed or
defective
– Check electrode and electrode
connection
– No calibration performed
– Calibrate
– Watering cap still on the
electrode
– Pull off watering cap and calibrate
– Electrode contaminated
– Clean electrode
– Liquid penetrated into the plug
connector
– Rinse the plug connector with
deionized water, clean and dry it
Cause
Remedy
– Battery depleted
– Insert new battery (see
Cause
Remedy
– Slope of the electrode not within
tolerance (see section 5.3)
– Condition the electrode
– If the slope is still outside the
tolerance, replace the electrode
– Check the buffer solutions
– Slope of the electrode too low
– Replace electrode
– Asymmetry of the electrode too
high
– Replace electrode