WQL-pH

What to do if...

141

ba75821e01

10/2009

12

What to do if...

12.1

Calibration and measuring

Lengthy stability

check with

AutoRead

Implausible

measured values

Logger does not

react

System cannot be

calibrated

Cause

Remedy

No stable measured value

Provide stable measuring conditions
(e.g. temperature)

Cause

Remedy

– Electrode not installed or

defective

– Check electrode and electrode

connection

– No calibration performed

– Calibrate

– Watering cap still on the

electrode

– Pull off watering cap and calibrate

– Electrode contaminated

– Clean electrode

– Liquid penetrated into the plug

connector

– Rinse the plug connector with

deionized water, clean and dry it

Cause

Remedy

– Battery depleted

– Insert new battery (see

section 11.4)

Cause

Remedy

– Slope of the electrode not within

tolerance (see section 5.3)

– Condition the electrode

– If the slope is still outside the

tolerance, replace the electrode

– Check the buffer solutions

– Slope of the electrode too low

– Replace electrode

– Asymmetry of the electrode too

high

– Replace electrode