Maintenance, cleaning, storage

WQL-pH

138

ba75821e01

10/2009

11.3

Replacing the electrode

Note
When replacing the electrode the general maintenance instructions in
section 11.1 must be followed.

The electrode is replaced in the same way as it was installed during the
commissioning (see section 3.1.3 I

NSTALLING

THE

ELECTRODE

).

11.4

Battery

11.4.1 Battery service life

The logger has an energy saving feature. The service life of the battery
depends very much on the usage of the logger, especially on the measuring
interval. The service life to be expected of a full battery (2600 mAh) can be
estimated by means of the following table:

Note
When inserting the battery, note down the installation time on the battery to
be able to estimate the remaining operational lifetime. When a battery is
inserted, the logger consumes a small amount of energy even when it is not
working. Therefore, we recommend to remove the battery during longer
measuring breaks.

11.4.2 Battery replacement

Note
When replacing the battery the general maintenance instructions in
section 11.1 must be followed.

The battery is replaced in the same way as it is inserted during the
commissioning (see section 3.1.2 I

NSERTING

THE

BATTERY

).

Measuring interval

Service time to be expected

1 sec

3 months

5 sec

5 months

10 sec

7 months

30 sec

11 months

1 min

13 months

5 min

15 months

10 min and longer

16 months