3 replacing the electrode, 4 battery, 1 battery service life – Xylem WQL-pH pH DATALOGGERS User Manual
Page 64: 2 battery replacement, Replacing the electrode, Battery 11.4.1, Battery service life, Battery replacement, Section 11.4 b, Attery

Maintenance, cleaning, storage
WQL-pH
138
ba75821e01
10/2009
11.3
Replacing the electrode
Note
When replacing the electrode the general maintenance instructions in
section 11.1 must be followed.
The electrode is replaced in the same way as it was installed during the
commissioning (see section 3.1.3 I
11.4
Battery
11.4.1 Battery service life
The logger has an energy saving feature. The service life of the battery
depends very much on the usage of the logger, especially on the measuring
interval. The service life to be expected of a full battery (2600 mAh) can be
estimated by means of the following table:
Note
When inserting the battery, note down the installation time on the battery to
be able to estimate the remaining operational lifetime. When a battery is
inserted, the logger consumes a small amount of energy even when it is not
working. Therefore, we recommend to remove the battery during longer
measuring breaks.
11.4.2 Battery replacement
Note
When replacing the battery the general maintenance instructions in
section 11.1 must be followed.
The battery is replaced in the same way as it is inserted during the
commissioning (see section 3.1.2 I
).
Measuring interval
Service time to be expected
1 sec
3 months
5 sec
5 months
10 sec
7 months
30 sec
11 months
1 min
13 months
5 min
15 months
10 min and longer
16 months