Silica lr vario, Si-1 tp (lr), 01 - 1.60 mg/l sio – YSI pHotoFlex Series User Manual
Page 60: 28 mm, Rn321

60
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:
Si-1 TP (LR)
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.01 - 1.60 mg/l SiO
2
0.005 - 0.748 mg/l Si
Display in mmol/l possible
Pipette 10.0 ml of sample
into the empty cell.
Add 15 drops of Vario
Molybdate 3 Reagent
Solution and close the
cell with the screw cap.
Mix the contents by
carefully swaying the
cell.
Allow to react for 4
minutes (temperature
dependence, see note).
Add the contents of a
VARIO Silica Citric
Acid F10 powder pack
and close the cell with
the screw cap.
Mix the contents by
carefully swaying the
cell.
Allow to react for 1
minute (temperature
dependence, see note).
Add the contents of a
VARIO LR Silica Amino
Acid F F10 powder
pack, close the cell with
the screw cap and mix.
Allow to react for 2
minutes. If SiO
2
is
present in the sample,
the solution becomes
blue.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
The above mentioned reaction times are valid at room temperature (20 °C). At 10 °C the reaction time has to be doubled,
at 30 °C it has to be halved.
1:00
2:00
Silica LR vario
Program no.