Chlorine (free) vario, Cl2-2 tp, 5 - 5.0 mg/l cl – YSI pHotoFlex Series User Manual
Page 24: 28 mm, Rn326

24
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:
Cl2-2 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.5 - 5.0 mg/l Cl
2
Display in mmol/l possible
Check the pH value of
the sample.
Desired range: pH 6-7.
Correct with diluted
sodium hydroxide
solution or sulfuric acid
as necessary.
Pipette 10.0 ml of
sample into an empty
beaker.
Add the contents of a
VARIO Chlorine FREE-
DPD F25 powder pack
and dissolve them by
stirring.
Add 15.0 ml deionized
water with a pipette and
mix.
Fill an empty cell with the
prepared sample and
close it with the screw
cap.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
The test sample should be pink. Very high chlorine concentrations in the sample cause yellow solutions and too low mea-
sured values. Dilute the sample in this case.
pH 6-7
Chlorine (free) vario
Program no.
326
*RN326*