Deha vario, Deha-1 tp, 004 - 0.450 mg/l deha – YSI pHotoFlex Series User Manual
Page 34: 28 mm, Rn335, 00 deha vario

34
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:
DEHA-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.004 - 0.450 mg/l DEHA
Display in mmol/l possible
Pipette 25.0 ml of
sample into an empty
beaker.
Add the contents of a
Vario Oxyscav 1 RGT
powder packet and
dissolve it by stirring.
Add 0.5 ml Vario DEHA
2 RGT with a pipette and
mix.
Fill an empty cell with the
prepared sample, close it
with the screw cap and
put it in a dark place.
Allow the sample to react
for ten minutes in a dark
place. Then measure
immediately.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
Clean all laboratory glassware with hydrochloric acid (approx. 20 %), then thoroughly rinse with deionized water.
•
Avoid excessive movements and exposure to sun light during sampling. Store the samples hermetically sealed.
•
The temperature of the samples must be 25±3 °C.
10:00
DEHA vario
Program no.
335
*RN335*