Ammonia (free), Mca-1 tp, 010 - 0.500 mg/l nh – YSI pHotoFlex Series User Manual
Page 16: N (f) 28 mm, Rn347

16
ba76126e01
03/2013
YSI model no.:
MCA-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.010 - 0.500 mg/l NH
3
-N (f)
0.012 - 0.610 mg/l NH
3
(f)
Display in mmol/l possible
Perform a zero
adjustment using a
28 mm cell filled with
sample.
Pipette 10.0 ml sample
into an empty 28 mm
cell.
Add 1 drop of Free
Ammonia Reagent
solution, close the cell
with the screw cap and
mix.
Allow to react for 5
minutes.
Add the contents of a
Vario Monochlor F
RGT powder pack and
close the cell with the
screw cap.
Note: Failing to restore
the zero adjustment
will result in all future
measurements being
erroneous!
Shake the cell vigorously
for about 20 seconds to
dissolve solids.
Allow to react for 5
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
When switching to
another program restore
the zero adjustment by
performing a new zero
adjustment using a
28 mm cell filled with
deionized water.
Notes:
•
Collect samples in clean glass bottles and analyze as soon as possible.
•
The 5 minutes reaction times refer to a sample temperature of 18-20 °C (64-68 °F).
For other sample temperatures adjust the reaction time according to the follwing table:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
5:00
5:00
Sample temperature
Reaction time
(minutes)
Sample temperature
Reaction time
(minutes)
°C
°F
°C
°F
5
41
10
20
68
5
10
50
8
23
73
2.5
16
61
6
25
77
2
Ammonia (free)
Program no.
347
*RN347*