Aluminium, Al-1 tp, 002 - 0.250 mg/l al – YSI pHotoFlex Series User Manual
Page 15: 28 mm, Rn323, 00 aluminium
15
ba76126e01
03/2013
Note: Before using this test with your photometer for the first time, determine the reagent blank
value (see notes below).
YSI model no.:
Al-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.002 - 0.250 mg/l Al
Pipette 20.0ml sample
into the empty cell.
Add the contents of a
VARIO Aluminum
ECR F20 powder pack
and close the cell with
the screw cap.
Dissolve the powder by
shaking.
Allow to react for 30
seconds.
Add the contents of a
VARIO Hexamine F20
powder pack and close
the cell with the screw
cap.
Dissolve the powder by
shaking.
For reagent blank value
only:
Add 2 drops of VARIO
ECR-Masking RGT,
close the cell with the
screw cap and mix.
Allow to react for 5
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
To avoid errors due to contamination, rinse the equipment with hydrochloric acid solution (approx. 20%), then deionized
water beforehand.
•
Adding the
VARIO ECR-Masking RGT
masking reagent is only required when determining the reagent blank value.
•
The sample temperature has to be between 20 and 25 °C.
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
5:00
Aluminium
Program no.
323
*RN323*