beautypg.com

Molybdate vario, Mo-1 tp, 3 - 35.0 mg/l mn – YSI pHotoFlex Series User Manual

Page 40: 28 mm, Rn304, 00 molybdate vario

background image

40

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

Mo-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.3 - 35.0 mg/l Mn

Display in mmol/l possible

Pipette 10.0 ml of sample
into the empty cell.

Add the contents of a
MolyVer 1 Reagenz
powder pack and close
the cell with the screw
cap.

Mix the contents by
carefully swaying the cell
(10 x).

Add the contents of a
MolyVer 2 Reagenz
powder pack and close
the cell with the screw
cap.

Mix the contents by
carefully swaying the cell
(10 x).

Add the contents of a
MolyVer 3 Reagenz
powder pack and close
the cell with the screw
cap.

Mix the contents by
carefully swaying the cell
(10 x).
Undissolved powder has
no adverse effect on
measurement.

Allow to react for 5
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

5:00

Molybdate vario

Program no.

304

*RN304*