Page 52: 28 mm, Rn306, 00 phosphate vario (ortho)

background image

52

ba76126e01

03/2013

Note: Before using the test with your photometer for the first time, determine the reagent blank
value.

YSI model no.:

PO4-1 TP

Category:

RT (reagent test)

Cell:

28 mm

Measuring range:

0.007 - 0.800 mg/l PO

4

-P

0.02 - 2.45 mg/l PO

4

Display in mmol/l possible

Pipette 10.0 ml of sample
into the empty cell.

Add the contents of a
VARIO Phos3 F10
powder pack and close
the cell with the screw
cap.

Shake the cell for 10 to
15 seconds.
Undissolved powder has
no adverse effect on
measurement.

Allow to react for 2
minutes.

Insert the cell in the
photometer cell shaft
and start measurement.

Notes:

We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.

2:00

Phosphate vario (ortho)

Program no.

306