Manganese vario, Mn-1 tp, 2 - 20.0 mg/l mn – YSI pHotoFlex Series User Manual
Page 38: 28 mm, Rn303, 00 manganese vario

38
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank
value.
YSI model no.:
Mn-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.2 - 20.0 mg/l Mn
Display in mmol/l possible
Check the pH value of
the sample.
Desired range: pH 4-5.
Correct with diluted nitric
acid or sodium hydroxide
solution as necessary.
Pipette 10.0 ml of sample
into the empty cell.
Add the contents of a
VARIO MANGANESE
Citrate Buffer F10
powder pack and close
the cell with the screw
cap.
Mix the contents by
carefully swaying the cell
(10 x).
Add the contents of a
VARIO Sodium
Periodate F10 powder
pack and close the cell
with the screw cap.
Mix the contents by
carefully swaying the cell
(10 x).
Allow to react for 2
minutes.
Insert the cell in the
photometer cell shaft
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
pH 4-5
2:00
Manganese vario
Program no.
303