Silica hr vario, Si-2 tp (hr), 7 - 70.0 mg/l sio – YSI pHotoFlex Series User Manual
Page 57: 28 mm, Rn307

57
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Si-2 TP (HR)
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.7 - 70.0 mg/l SiO
2
0.3 - 32.7 mg/l Si
Display in mmol/l possible
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
VARIO Silica HR 
Molybdate F10 powder 
pack and close the cell 
with the screw cap.
Mix the contents by 
carefully swaying the 
cell.
Add the contents of a 
VARIO Silica HR Acid 
Rgt F10 powder pack 
and close the cell with 
the screw cap.
Mix the contents by 
carefully swaying the 
cell.
Allow to react for 10 
minutes.
Add the contents of a 
VARIO Silica Citric 
Acid F10 powder pack 
and close the cell with 
the screw cap.
Mix the contents by 
carefully swaying the 
cell.
Allow to react for 2 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
The sample temperature has to be between 15 and 25 °C.
10:00
2:00
Silica HR vario
Program no.
307
*RN307*