Nitrite lr, No2-2 tc, 03 - 0.60 mg/l no – YSI pHotoFlex Series User Manual
Page 46: 16 mm, Rn318, 00 nitrite lr

46
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
NO2-2 TC
Category:
KT (reaction cell test)
Cell:
16 mm
Measuring range:
0.03 - 0.60 mg/l NO
2
-N
0.10 - 1.97 mg/l NO
2
Display in mmol/l possible
Pipette 2.0 ml of sample 
into a reaction cell.
Mix the contents by 
carefully swaying the 
cell.
Add 1 level black 
measuring spoon no. 8 of 
Nitrit-101 and close the 
cell with the screw cap.
Shake the cell vigorously 
to dissolve solids.
Allow to react for 10 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
Store the closed reagents at a temperature of +4 to +8 °C.
10:00
Nitrite LR
Program no.
318
*RN318*
- 3100 (50 pages)
 - 3200 (86 pages)
 - 5000 (70 pages)
 - 5200 (148 pages)
 - 5200A (121 pages)
 - 5200A (128 pages)
 - 5400 (127 pages)
 - 5239 Dissolved Oxygen Probe (23 pages)
 - 55 (26 pages)
 - 550A (18 pages)
 - 556 MPS - Quick Start Guide (4 pages)
 - 58 (40 pages)
 - 5905 (22 pages)
 - 6600V2 - User Manual (379 pages)
 - 60 (33 pages)
 - 63 (46 pages)
 - 650 MDS (72 pages)
 - 9300 Photometer (11 pages)
 - 9500 Photometer (143 pages)
 - 9500 Photometer (18 pages)
 - CR 3200 Thermoreactor (42 pages)
 - CR 4200 Thermoreactor (42 pages)
 - CR2200 Thermoreactor (36 pages)
 - DO200 (40 pages)
 - EC300 (40 pages)
 - EC300A (40 pages)
 - EC30A (12 pages)
 - IFL 700 IQ (46 pages)
 - IQ SensorNet 2020 XT Terminal (196 pages)
 - IQ SensorNet FDO Sensor (48 pages)
 - IQ SensorNet MIQ/A-MOD Module (42 pages)
 - IQ SensorNet MIQ/A-PR Module (36 pages)
 - IQ SensorNet MIQ/IC2 Module (30 pages)
 - IQ SensorNet MIQ/IF232 Module (18 pages)
 - IQ SensorNet MIQ/JB Module (18 pages)
 - IQ SensorNet MIQ/JBR Module (20 pages)
 - IQ SensorNet MIQ/PS Module (22 pages)
 - IQ SensorNet MIQ/R6 Module (56 pages)
 - IQ SensorNet NitraLyt Sensor (64 pages)
 - IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
 - IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
 - IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
 - IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
 - MultiLab 4010-1 (108 pages)
 - MultiLab 4010-2 (138 pages)
 