Nitrate, No3-1 tc, 2 - 30.0 mg/l no – YSI pHotoFlex Series User Manual
Page 44: 16 mm, Rn314, 00 nitrate

44
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
NO3-1 TC
Category:
KT (reaction cell test)
Cell:
16 mm
Measuring range:
0.2 - 30.0 mg/l NO
3
-N
1.0 - 133.0 mg/l NO
3
Display in mmol/l possible
Pipette 1.0 ml of sample 
into a reaction cell and 
close the cell with the 
screw cap.
Mix the contents by 
carefully swaying the cell 
(10 x).
Add the contents of a 
VARIO Nitrate 
Chromotropic powder 
pack and close the cell 
with the screw cap.
Mix the contents by 
carefully swaying the cell 
(10 x). A small amount of 
solids may remain 
undissolved.
Allow to react for 5 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
5:00
Nitrate
Program no.
314
*RN314*
- 3100 (50 pages)
 - 3200 (86 pages)
 - 5000 (70 pages)
 - 5200 (148 pages)
 - 5200A (128 pages)
 - 5200A (121 pages)
 - 5400 (127 pages)
 - 5239 Dissolved Oxygen Probe (23 pages)
 - 55 (26 pages)
 - 550A (18 pages)
 - 556 MPS - Quick Start Guide (4 pages)
 - 58 (40 pages)
 - 5905 (22 pages)
 - 6600V2 - User Manual (379 pages)
 - 60 (33 pages)
 - 63 (46 pages)
 - 650 MDS (72 pages)
 - 9500 Photometer (143 pages)
 - 9300 Photometer (11 pages)
 - 9500 Photometer (18 pages)
 - CR 3200 Thermoreactor (42 pages)
 - CR 4200 Thermoreactor (42 pages)
 - CR2200 Thermoreactor (36 pages)
 - DO200 (40 pages)
 - EC300 (40 pages)
 - EC300A (40 pages)
 - EC30A (12 pages)
 - IFL 700 IQ (46 pages)
 - IQ SensorNet 2020 XT Terminal (196 pages)
 - IQ SensorNet FDO Sensor (48 pages)
 - IQ SensorNet MIQ/A-MOD Module (42 pages)
 - IQ SensorNet MIQ/A-PR Module (36 pages)
 - IQ SensorNet MIQ/IC2 Module (30 pages)
 - IQ SensorNet MIQ/IF232 Module (18 pages)
 - IQ SensorNet MIQ/JB Module (18 pages)
 - IQ SensorNet MIQ/JBR Module (20 pages)
 - IQ SensorNet MIQ/PS Module (22 pages)
 - IQ SensorNet MIQ/R6 Module (56 pages)
 - IQ SensorNet NitraLyt Sensor (64 pages)
 - IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
 - IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
 - IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
 - IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
 - MultiLab 4010-1 (108 pages)
 - MultiLab 4010-2 (138 pages)
 