Molybdate vario, Mo-1 tp, 3 - 35.0 mg/l mn – YSI pHotoFlex Series User Manual
Page 40: 28 mm, Rn304, 00 molybdate vario

40
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Mo-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.3 - 35.0 mg/l Mn
Display in mmol/l possible
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
MolyVer 1 Reagenz 
powder pack and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the cell 
(10 x).
Add the contents of a 
MolyVer 2 Reagenz 
powder pack and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the cell 
(10 x).
Add the contents of a 
MolyVer 3 Reagenz 
powder pack and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the cell 
(10 x). 
Undissolved powder has 
no adverse effect on 
measurement.
Allow to react for 5 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
5:00
Molybdate vario
Program no.
304
*RN304*