Manganese vario, Mn-2 tp, 007 - 0.700 mg/l mn – YSI pHotoFlex Series User Manual
Page 39: 28 mm, Rn330, 00 manganese vario

39
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Mn-2 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.007 - 0.700 mg/l Mn
Display in mmol/l possible
Check the pH value of 
the sample. 
Required range: pH 4-5. 
Correct with diluted nitric 
acid or sodium hydroxide 
solution as necessary.
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
Vario Ascorbic Acid 
powder packet and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the 
cell.
Add 15 drops of Vario 
Alkaline-Cyanide 
Reagent Solution and 
close the cell with the 
screw cap.
Mix the contents by 
carefully swaying the 
cell.
Add 21 drops of Vario 
PAN Indicator Solution 
0.1% and close the cell 
with the screw cap.
Mix the contents by 
carefully swaying the 
cell.
Allow to react for 2 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
Clean all laboratory glassware with nitric acid, then thoroughly rinse with deionized water.
pH 4-5
2:00
Manganese vario
Program no.
330
*RN330*