Manganese vario, Mn-1 tp, 2 - 20.0 mg/l mn – YSI pHotoFlex Series User Manual
Page 38: 28 mm, Rn303, 00 manganese vario

38
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Mn-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.2 - 20.0 mg/l Mn
Display in mmol/l possible
Check the pH value of 
the sample. 
Desired range: pH 4-5. 
Correct with diluted nitric 
acid or sodium hydroxide 
solution as necessary.
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
VARIO MANGANESE 
Citrate Buffer F10 
powder pack and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the cell 
(10 x).
Add the contents of a 
VARIO Sodium 
Periodate F10 powder 
pack and close the cell 
with the screw cap.
Mix the contents by 
carefully swaying the cell 
(10 x).
Allow to react for 2 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
pH 4-5
2:00
Manganese vario
Program no.
303
*RN303*
