Iron vario, Fe-2 tp, 02 - 3.00 mg/l fe – YSI pHotoFlex Series User Manual
Page 36: 28 mm, Rn301, 00 iron vario

36
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Fe-2 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.02 - 3.00 mg/l Fe
Display in mmol/l possible
Check the pH value of 
the sample. 
Desired range: pH 3-5. 
Correct with diluted 
sodium hydroxide 
solution or hydrochloric 
acid as necessary.
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
VARIO Ferro F10 
powder pack and close 
the cell with the screw 
cap.
Mix the contents by 
carefully swaying the cell 
(10 x). 
Undissolved powder has 
no adverse effect on 
measurement.
Allow to react for 3 
minutes (reaction time).
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
This method covers all forms of dissolved iron and most forms of undissolved iron.
•
In the case of samples with visible rust the reaction time should be extended to at least 5 minutes.
pH 3-5
3:00
Iron vario
Program no.
301
*RN301*