Copper vario, Cu-1 tp, 04 - 5.00 mg/l cu – YSI pHotoFlex Series User Manual
Page 33: 28 mm, Rn302, 00 copper vario

33
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Cu-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.04 - 5.00 mg/l Cu
Display in mmol/l possible
Check the pH value of 
the sample. 
Desired range: pH 4-6. 
Correct with diluted 
sodium hydroxide 
solution or caustic 
potash solution as 
necessary.
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
VARIO Cu1 F10 powder 
pack and close the cell 
with the screw cap.
Mix the contents by 
carefully swaying the cell 
(10 x). 
Undissolved powder has 
no adverse effect on 
measurement.
Allow to react for 2 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
pH 4-6
2:00
Copper vario
Program no.
302
*RN302*
- 3100 (50 pages)
 - 3200 (86 pages)
 - 5000 (70 pages)
 - 5200 (148 pages)
 - 5200A (121 pages)
 - 5200A (128 pages)
 - 5400 (127 pages)
 - 5239 Dissolved Oxygen Probe (23 pages)
 - 55 (26 pages)
 - 550A (18 pages)
 - 556 MPS - Quick Start Guide (4 pages)
 - 58 (40 pages)
 - 5905 (22 pages)
 - 6600V2 - User Manual (379 pages)
 - 60 (33 pages)
 - 63 (46 pages)
 - 650 MDS (72 pages)
 - 9300 Photometer (11 pages)
 - 9500 Photometer (143 pages)
 - 9500 Photometer (18 pages)
 - CR 3200 Thermoreactor (42 pages)
 - CR 4200 Thermoreactor (42 pages)
 - CR2200 Thermoreactor (36 pages)
 - DO200 (40 pages)
 - EC300 (40 pages)
 - EC300A (40 pages)
 - EC30A (12 pages)
 - IFL 700 IQ (46 pages)
 - IQ SensorNet 2020 XT Terminal (196 pages)
 - IQ SensorNet FDO Sensor (48 pages)
 - IQ SensorNet MIQ/A-MOD Module (42 pages)
 - IQ SensorNet MIQ/A-PR Module (36 pages)
 - IQ SensorNet MIQ/IC2 Module (30 pages)
 - IQ SensorNet MIQ/IF232 Module (18 pages)
 - IQ SensorNet MIQ/JB Module (18 pages)
 - IQ SensorNet MIQ/JBR Module (20 pages)
 - IQ SensorNet MIQ/PS Module (22 pages)
 - IQ SensorNet MIQ/R6 Module (56 pages)
 - IQ SensorNet NitraLyt Sensor (64 pages)
 - IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
 - IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
 - IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
 - IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
 - MultiLab 4010-1 (108 pages)
 - MultiLab 4010-2 (138 pages)
 