Chlorine (free) vario, Cl2-2 tp, 5 - 5.0 mg/l cl – YSI pHotoFlex Series User Manual
Page 24: 28 mm, Rn326

24
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
Cl2-2 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
0.5 - 5.0 mg/l Cl
2
Display in mmol/l possible
Check the pH value of 
the sample. 
Desired range: pH 6-7. 
Correct with diluted 
sodium hydroxide 
solution or sulfuric acid 
as necessary.
Pipette 10.0 ml of 
sample into an empty 
beaker.
Add the contents of a 
VARIO Chlorine FREE-
DPD F25 powder pack 
and dissolve them by 
stirring.
Add 15.0 ml deionized 
water with a pipette and 
mix.
Fill an empty cell with the 
prepared sample and 
close it with the screw 
cap.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
The test sample should be pink. Very high chlorine concentrations in the sample cause yellow solutions and too low mea-
sured values. Dilute the sample in this case.
pH 6-7
Chlorine (free) vario
Program no.
326
*RN326*
