Ammonium vario hr, 4 - 50.0 mg/l nh, Nh4-3 tc (hr) – YSI pHotoFlex Series User Manual
Page 21: 16 mm, Rn313, 00 ammonium vario hr

21
ba76126e01
03/2013
Note: Before using the test with your photometer for the first time, determine the reagent blank 
value.
YSI model no.:
NH4-3 TC (HR)
Category:
KT (reaction cell test)
Cell:
16 mm
Measuring range:
0.4 - 50.0 mg/l NH
4
-N
0.5 - 64.4 mg/l NH
4
Display in mmol/l possible
Check the pH value of 
the sample. 
Desired value: approx. 
pH 7. 
Correct with diluted 
sodium hydroxide 
solution or hydrochloric 
acid as necessary.
Pipette 0.1 ml of sample 
into a reaction cell.
Add the contents of a 
VARIO AMMONIA 
Salicylate F5 powder 
pack.
Add the contents of a 
VARIO AMMONIA 
Cyanurate F5 powder 
pack and close the cell 
with the screw cap.
Shake the cell vigorously 
to dissolve solids.
Allow to react for 20 
minutes.
Insert the cell in the 
photometer cell shaft 
and start measurement.
Notes:
•
We recommend to determine a new reagent blank value (deionized water instead of sample) for each test package started.
•
If NH
4
-N is present in the sample, the solution becomes green after VARIO AMMONIA Cyanurate F5 was added.
•
If chlorine is present, sodium thiosulfate has to be added to the sample immediately after sampling. Add 1 drop of a 0.1 
mol/l sodium thiosulfate solution per 0.3 mg/l chlorine to 1 liter sample.
•
Iron disturbs the measurement and can be eliminated as follows: Determine the total iron concentration and prepare an 
iron standard solution with the determined concentration. Use this solution instead of distilled water to determine the 
reagent blank value for ammonium measurement.
approx. 
pH 7
20:00
Ammonium vario HR
Program no.
313
*RN313*