Ammonia, Nh4-1 tp, 28 mm – YSI pHotoFlex Series User Manual
Page 17: Rn341

17
ba76126e01
03/2013
Note: Before using tYSIhe test with your photometer for the first time, determine the reagent 
blank value. To determine the reagent blank value it is not necessary to measure the pH and 
temperature.
Step 1: pH and temperature measurement
Step 2: photometric measurement
YSI model no.:
NH4-1 TP
Category:
RT (reagent test)
Cell:
28 mm
Measuring range:
Corresponding to 0.01 - 0.64 mg/l NH
4
or 0.01 - 0.50 mg/l NH
4
-N
Measuring ranges for NH
3
or NH
3
-N depending on pH value and temperature,
Example: 0.002 - 0.092 mg/l NH
3
at pH 8.5 and 25 °C.
Measuring mode, pH & 
ORP:
Measure pH value and 
temperature immediately 
after sampling.
Store measured values 
with 
sary, assign an ID for 
easier retrieving.
Switch to the Photometry 
measuring mode and 
select program no. 341.
When the prompt Assign 
pH und temp.
appears,
select and accept the 
stored values from the 
pH and temperature 
measurement.
The meter is ready for 
the photometric mea-
surement (step 2). 
The measuring range is 
shown on the display.
Pipette 10.0 ml of sample 
into the empty cell.
Add the contents of a 
VARIO AMMONIA 
Salicylate F10 powder 
pack and close the cell 
with the screw cap.
Allow to react for 3 
minutes (reaction time).
Add the contents of a 
VARIO AMMONIA 
Cyanurate F10 powder 
pack and close the cell 
with the screw cap.
Shake the cell vigorously 
to dissolve solids.
Allow to react for 15 
minutes (reaction time).
Insert the cell in the 
photometer cell shaft 
and start measurement.
3:00
15:00
Notes:
•
The measuring range depends largely on the pH and temperature. On 
the basis of the pH and temperature measurement, it is individually cal-
culated and displayed for each determination.
•
We recommend to determine a new reagent blank value (deionized 
water instead of sample) for each test package started. To determine 
the reagent blank value it is not necessary to measure the pH and tem-
perature.
•
For further notes please refer to the package insert of the test.
Ammonia
Program no.
341