Storing user-defined
programs
Enter a command line according to the following scheme in the terminal
program:
U.500#14,Program No,Model,Cell [mm],Wave length [nm],MRB,
MRE,E0,Slope,Resolution,Citation form,Unit,Blank required,Standard
adjustment possible,Standard adjustment required
Example:
U.500#14,900,Test,16,436,0.0,2.0,0.0,1.0,0.01,test,
mg/l,0,1,0
The individual data sections of the command line are separated by
commas. The dot "." has to be used as a decimal separator within a
data section.
If a program is already stored under the selected number, the program-
ming procedure is canceled. To store the program under the selected
number, first erase the program stored under the number.
After successful transmission, the terminal program writes "!>".
If the transmission failed, the terminal program writes "!>".
Erasing user-defined
programs
To erase user-defined programs, enter a command line according to
the following scheme in the terminal program:
After successful transmission, the requested data is displayed on the
terminal.
If the transmission failed, the terminal program writes "!>".
5
Enter the command line.
6
Finish the command line with Enter.
The data is transmitted to the pHotoFlex
®
pH.
Erase all user-defined
programs
Erase one user-
defined program
Command line
U.520
U.521#1,Program No
Example
U.520
U.521#1,900
7
Enter the command line.
8
Finish the command line with Enter.
The data is transmitted to the pHotoFlex
®
pH.
The requested data is displayed on the terminal as the result.