beautypg.com

2 measuring the ph value, Measuring the ph value – YSI pHotoFlex pH Handheld Colorimeter User Manual

Page 50

background image

Operation

pHotoFlex

®

pH

50

ba76123e01

02/2013

4.6.2

Measuring the pH value

AutoRead

(Drift control)

The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured pa-
rameter flashes until a stable measured value is available.

Criteria

With identical measurement conditions, the following applies:

1

Perform the preparatory activities according to section 4.6.1.

2

Immerse the pH electrode in the test sample.

3

Press the key (short pressure) repeatedly until the mea-
sured variable, pH is selected.

pH & ORP \ pH

6.94

25.0

°C

[AR]

01.02.05 15:12

Measured
parameter

Reproducibility

Response time

pH value

Better than 0.01

> 30 seconds