Measuring the ORP voltage
ORP electrodes are not calibrated. However, you can check ORP elec-
trodes using a test solution.
AutoRead
(drift control)
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured pa-
rameter flashes until a stable measured value is available.
Criteria
With identical measurement conditions, the following applies:
4.6.4
Settings for pH and ORP measurements
Overview
For pH and ORP measurements, the following settings are available in
the Configuration / pH & ORP menu:
Measured parameter
Calibration record (display)
Selecting the calibration type
Entering the Calibration interval
Selecting the Unit for slope
Selecting the Temperature unit
Reset
1
Perform the preparatory activities according to section 4.6.1.
2
Submerse the ORP electrode in the sample.
3
Press the
sured parameter, ORP is selected.
pH & ORP \ ORP
157
mV
25.0
°C
[AR]
01.02.05 15:12
Measured
parameter
Reproducibility
Response time
ORP voltage
better than 1 mV
> 30 seconds